: Instrument Messages : Self-Test Messages
 
Self-Test Messages
The MG362x1A signal generator has built-in diagnostics that allow you to run a built-in self test. The test results can be viewed on the screen and saved to a log file. To run self test, use the 3-Line Icon to access DIAGNOSTICS, then select SELF TEST and RUN SELF TEST. Below is a sample list of self tests with passing results:
 System Messages
Name
Type
Actual
Expected
Tolerance
Status
System voltage
Voltage
13.02 V
13.00 V
4.00 V
pass
+1.0V digital supply
Voltage
0.99 V
1.00 V
0.10 V
pass
+3.3V digital supply
Voltage
3.28 V
3.30 V
0.33 V
pass
+5.0V digital supply
Voltage
5.00 V
5.00 V
0.50 V
pass
+1.8V digital supply
Voltage
1.83 V
1.80 V
0.18 V
pass
+0.85V PS transceiver AVCC
Voltage
0.84 V
0.85 V
0.09 V
pass
+1.8V PS transceiver AVTT
Voltage
1.80 V
1.80 V
0.18 V
pass
+2.5V DDR SODIMM supply
Voltage
2.49 V
2.50 V
0.25 V
pass
+1.2V DDR SODIMM supply
Voltage
1.20 V
1.20 V
0.12 V
pass
+0.85V FPGA core supply
Voltage
0.85 V
0.85 V
0.09 V
pass
CPU Temperature
Temperature
37.40 C
50.00 C
50.00 C
pass
 
 SG Messages
Name
Type
Actual
Expected
Tolerance
Status
MOD +2.5V supply
Voltage
2.49 V
2.50 V
0.25 V
pass
MOD +3.3V supply
Voltage
3.31 V
3.30 V
0.33 V
pass
ALC
Temperature
59.56 C
30.00 C
60.00 C
pass
YIG
Temperature
40.25 C
30.00 C
60.00 C
pass
SWF +2.5V supply
Voltage
2.48 V
2.50 V
0.25 V
pass
SWF +3.3V supply
Voltage
3.33 V
3.30 V
0.33 V
pass
SWF +5.0V supply
Voltage
5.02 V
5.00 V
0.50 V
pass
SWF +8.0V supply
Voltage
8.07 V
8.00 V
0.80 V
pass
SWF +10.0V HP supply
Voltage
9.65 V
10.00 V
1.00 V
pass
SWF +28.0V supply
Voltage
27.68 V
28.00 V
2.80 V
pass
SWF -2.5V HP supply
Voltage
-2.52 V
-2.50 V
0.25 V
pass
SWF -5.0V HP supply
Voltage
-5.06V
-5.00 V
0.50 V
pass
SWF -32.0V HP supply
Voltage
-31.80 V
-32.00 V
3.20 V
pass
SWF -37.0V HP supply
Voltage
-36.41 V
-37.00 V
3.70 V
pass
SWF -3.3V HP supply
Voltage
-3.38 V
-3.30 V
0.33 V
pass
SWF ALC1 current monitor
Current
0.05 A
0.40 A
0.40 A
pass
SWF ALC2 current monitor
Current
0.32 A
0.40 A
0.40 A
pass
SWF DA1 current monitor
Current
0.41 A
0.40 A
0.40 A
pass
SWF DA2 current monitor
Current
0.00 A
0.40 A
0.40 A
pass
SWF PA1 current monitor
Current
0.30 A
0.70 A
0.70 A
pass
SWF PA2 current monitor
Current
0.25 A
0.70 A
0.70 A
pass
SWF PA3 current monitor
Current
0.50 A
0.70 A
0.70 A
pass
SWF Temperature
Temperature
32.00 C
30.00 C
60.00 C
pass
Frequency Extension Module Not Available in MG36221A
Voltage
0.00 V
0.00 V
0.00 V
pass
REG +13V Main Voltage
Voltage
13.04 V
13.00 V
1.30 V
pass
REG +6.0V Standby
Voltage
6.01 V
6.00 V
0.60 V
pass
REG +1.5V supply
Voltage
1.52 V
1.50 V
0.15 V
pass
REG +3.3V supply
Voltage
3.35 V
3.30 V
0.33 V
pass
REG +3.9V supply
Voltage
4.03 V
3.90 V
0.39 V
pass
REG +6.0V supply
Voltage
6.02 V
6.00 V
0.60 V
pass
REG +9.0V supply
Voltage
9.18 V
9.00 V
0.90 V
pass
REG +10.0V supply
Voltage
10.18 V
10.00 V
1.00 V
pass
REG +15.0V supply
Voltage
15.10 V
15.00 V
1.50 V
pass
REG +16.0V supply
Voltage
16.59 V
16.00 V
1.60 V
pass
REG +24.0V supply
Voltage
24.22 V
24.00 V
2.40 V
pass
REG +25.0V supply
Voltage
26.24 V
25.00 V
2.50 V
pass
REG +30.0V supply
Voltage
30.13 V
30.00 V
3.00 V
pass
REG -6.0V supply
Voltage
-5.86 V
-6.00 V
0.60 V
pass
REG -16.0V supply
Voltage
-15.76 V
-16.00 V
1.60 V
pass
REG -15.0V supply
Voltage
-14.61 V
-15.00 V
1.50 V
pass
REG -30.0V supply
Voltage
-28.59 V
-30.00 V
3.00 V
pass
REG -40.0V supply
Voltage
-38.39 V
-40.00 V
4.00 V
pass
REG -6.0V standby
Voltage
-5.81 V
-6.00 V
0.60 V
pass
REG -13.0V standby
Voltage
-12.57 V
-13.00 V
1.30 V
pass
Overall Status
Info value
 
pass