Remote Spectrum Monitor™ MS27201A User Guide : 5GNR Measurements : 5GNR Measurement Setup
 
5GNR Measurement Setup
5GNR base station validation qualifies the performance of gNB base stations with essential measurements that are in full compliance with 3GPP TS 38.104 V15. The 5GNR analyzer features smart measurements that make configuring these measurements and views for 5GNR compliance testing easy:
5GNR summary with multi beam and single beam views measures:
Frequency Error
Time Offset (time difference between the measured start of the 5GNR frame and the 10 ms frame trigger, which is automatically synchronized to GPS when the instrument has GPS lock)
Cell/Sector ID
Modulation Quality
Synchronization Signal Block (SSB)
Error Vector Magnitude (EVM, peak and average)
Secondary Synchronization Reference Signal Received Power (SS-RSRP)
Secondary Synchronization Signal Reference Signal Received Quality (SS-RSRQ)
Secondary Synchronization Signal-to-noise and Interference Ratio (SS-SINR)
5GNR Multi PCI with multi beam and table views enhances the summary measurements with the ability to simultaneously measure multiple physical-layer cells.
Occupied Bandwidth (OBW)
Channel Power (CP)
Equivalent Isotropic Radiated Power (EIRP)
Carrier Aggregation
Constellation
MEASURE Menu (5GNR)
 
5GNR MEASURE Menu
-
 
MEASUREMENT
Selects the desired measurement type from the following list:
5GNR Summary: Displays the current beam measurement status in a quick-view summary that includes the PCI summary, sync signal (SS), and channel summary. See 5GNR Summary and Multi PCI.
Channel Power: Provides an RF channel power measurement that includes the total channel power, power spectral density (PSD), and pass/fail compliance testing. 5GNR Channel Power.
OBW: Provides an RF occupied bandwidth measurement that includes the occupied bandwidth, total power, and pass/fail compliance testing. 5GNR Occupied Bandwidth.
EIRP: Provides an EIRP display that includes the input parameters setup and measurement status. 5GNR EIRP.
Multi PCI: Displays the current beam measurement status of multiple cells in a quick-view summary that includes the PCI summary, sync signal (SS), and channel summary. See 5GNR Summary and Multi PCI.
Carrier Aggregation: Provides a component carrier measurement that includes the PCI, reference signal power, EVM, and frequency error measurement summaries. See 5GNR Carrier Aggregation.
Constellation: Provides a constellation measurement that includes the PCI summary, reference signal power, EVM, and frequency error measurement summaries. See 5GNR Constellation.
VIEW
When the measurement is 5GNR Summary, you can select the type of view:
Multi Beam: Displays eight beam power levels and measurement summaries on a total of eight summary pages. See 5GNR Summary Multi Beam View.
Single Beam: Displays a total of 64 individual summary views for each beam. See 5GNR Summary Single Beam View.
When the measurement is 5GNR Multi PCI, you can select the type of view:
Scanner: Displays eight beam power levels and summaries on a total of eight summary pages. See 5GNR Multi PCI Scanner View.
Table: Displays all received PCI measurement data in a single table view. See 5GNR Multi PCI Table View.
SETUP
Opens one of the measurement SETUP menus. The 5GNR SETUP menus can be access directly via the SETUP or MEASURE menus after the desired measurement has been selected. The setup menu depends on the current measurement and view selected.