Description | Set or return the duty cycle correction factor currently in use on the selected channel. |
Syntax: | CALCulate[1|2|3|4]:CORRection:DCYCle <numeric value> |
Argument: | <numeric_value> = 0.01 to 100.00 percent Valid Modes: CW Sensor (Modulated Mode) |
Modes: | Any |
Description | Set or return the system measurement mode. MODULATED mode is a continuous measurement mode primarily for continuously modulated or CW signals. PULSE mode is a signal triggered, oscilloscope-like mode that acquires and analyzes a pulsed signal as a series of one or more triggered sweeps. STATISTICAL mode performs long-term power distribution analysis on modulated signals, and may be operated in a start-stop continuous mode, or a decimated continuous mode. Note that the measurement mode is global and affects all channels. If any channel has a CW sensor installed, peak or statistical measurements will be unavailable, but the “primary “average power measurement will still be performed. |
Syntax: | CALCulate:MODE <character data> |
Argument: | <character data> = {MODulated, PULSe, STATistical} |
Modes: | Any |
Description: | Set or return the operating mode of the selected channel’s peak hold function. When set to OFF, instantaneous peaks are only held for a short time, and then decayed towards the average power at a rate proportional to the filter time. This is the best setting for most signals, because the peak will always represent the peak power of the current signal, and the resulting peak-to-average ratio will be correct shortly after any signal level changes. When set to ON, instantaneous peaks are held until reset by a new INITiate command or cleared manually. This setting is used when it is desirable to hold the highest peak over a long measurement interval without any decay. |
Syntax: | CALCulate[1|2|3|4]:PKHLD <Boolean> |
Argument: | <Boolean> = { 0, 1, OFF, AVG, INST } Valid Modes: Modulated and Pulse Modes |
Description: | Set or return the measurement state of the selected channel. When ON, the channel performs measurements; when OFF, the channel is disabled and no measurements are performed. |
Syntax: | CALCulate[1|2|3|4]:STATe <Boolean> Argument: <Boolean> = { 0, 1, OFF, ON } |
Valid Modes: | Any |
Description: | Set or return units for the selected channel. For power sensors, voltage is calculated with reference to the sensor input impedance. Note that for ratiometric results, logarithmic units will always return dBr (dB relative) while linear units return percent. |
Syntax: | CALCulate[1|2|3|4]:UNITs <character data> |
Argument: | <character data> = {DBMw, Watts, Volts, DBV, DBMV, DBUV} |
Valid Modes: | Any |