VectorStar™ MS4640A Series Microwave Vector Network Analyzers : Calibration Overview : SOLT Kits (3650x and 3654x)
 
SOLT Kits (3650x and 3654x)
These kits, all based on short-open-load-through, require data describing all of the reflection standards (provided by the factory) be loaded into the instrument on a serial number basis. If this media (a USB key) is not available, average default coefficients are available within the VNA that may suffice for some measurements.
Typically these calibration kits are loaded using the CAL KIT/AUTOCAL utility menu (Figure: CAL KIT/AUTOCAL Utility Menu) but user-defined kits can also be created using the parameters described above. If calibration kits from another manufacturer are used or to create a calibration kit, the following parameters are typically entered into one of the user-defined kits:
Open definition (M and F typically)
Short definition (M and F typically)
Load definition (M and F typically)
 
CAL KIT/AUTOCAL Utility Menu
The CAL KIT/AUTOCAL utility menu is shown here. This menu allows calibration kit details to be loaded from external files (as provided with Anritsu calibration kits), saved to a file (for user-defined cal kits), defaulted (for standard connector types), or simply displayed.