VectorStarâ„¢ MS4640A Series Microwave Vector Network Analyzers : Other Calibration Procedures : Through (Thru) Update
 
Through (Thru) Update
A common question related to calibrations is how often the calibration must be redone. The frequency of re-calibration depends on the environment (both in terms of temperature stability and in terms of the cable/fixture construct that is being used). A calibration lifetime is often limited by the stability of the test port cables through drift or motion. Drift defects directly affect transmission tracking and load match, so if those terms could be easily refreshed, the time before the entire calibration had to be redone could be extended. This is the concept of a Thru Update.
The idea is to connect just a through line and quickly refresh the transmission tracking and load match terms without great effort. A Thru Update is a one-step calibration that can be used to refresh a current full 2-Port or 1 path-2 port calibration.
Like the thru step in other calibrations, the length of the line being used as well as its loss can be specified to ensure minimal disruption of the reference planes (see Figure: THRU UPDATE Setup Menu and THRU INFO Dialog). The entry methods for these parameters are the same as in the normal calibration procedures.