VectorStar™ MS4640A Series Microwave Vector Network Analyzers : Mixer Setup and Measurement : Procedure for a Typical Single Channel Mixer Measurement
 
Procedure for a Typical Single Channel Mixer Measurement
Consider that one wishes to make a conversion loss measurement of a DUT with a 10 to 20 GHz input range, a fixed 0.5 GHz output and a low-side LO. External synthesizer 1 will be used as the LO (and it is on the GPIB address of 4 which is default for external synth 1) and it has been connected to the GPIB and the 10 MHz reference out of the VNA and turned on. The LO power is to be +10 dBm and the input power is to be -10 dBm. The setup dialogs for this scenario are shown in Figure: Example Setup for 10 to 20 GHz Downconverter. Here we used the ‘auto’ function on the LO entry to simplify things. Note the selection of the conversion direction on the frequency setup. Once the Done button is selected on the main dialog, the setup will be verified and the system placed into a converting mode where the source at port 1 is sweeping over 10 to 20 GHz and the receiver is fixed at 0.5 GHz.
Example Setup for 10 to 20 GHz Downconverter
An example setup is shown here for a 10 to 20 GHz downconverter with a fixed IF of 0.5 GHz. The LO range will be calculated to be 9.5 to 19.5 GHz and programmed into an external synthesizer.
 
The next step would normally be to go to the calibration menu and select either a normalization or an enhanced match calibration for this conversion loss measurement. During the normalization calibration, only a thru line connection is needed as discussed earlier. The measurement result for this case is shown in Figure: Example Conversion Gain Measurement with Normalization-Based Calibration. No port padding was used and the DUT was not particularly well-matched so there is some ripple.
Example Conversion Gain Measurement with Normalization-Based Calibration
An example conversion gain measurement is shown here using the setup of Figure: Example Setup for 10 to 20 GHz Downconverter above and a normalization-based calibration.
 
An enhanced-match calibration was also performed on this setup and the measurement results are shown in Figure: Example Conversion Gain Measurement with Enhanced Match Calibration. A standard manual OSL calibration kit in the K connector (3652X) was used for the reflectometer calibration steps. The measurements were not identical due to some bias changes on the DUT but the reduction in ripple is apparent.
Example Conversion Gain Measurement with Enhanced Match Calibration
An example conversion gain measurement is shown here using the setup of Figure: Example Setup for 10 to 20 GHz Downconverter and an enhanced match calibration.