VectorStar™ MS4640A SeriesMicrowave Vector Network Analyzer : SCPI Commands : :CALCulate{1-16}[:SELected]:GCOMpression Subsystem
 
:CALCulate{1-16}[:SELected]:GCOMpression Subsystem
The :CALCulate{1-16}[:SELected]:GCOMpression subsystem configures and controls the gain compression instrument functions.
Time Domain, Group Delay, and Reference Plane Subsystems
Related time domain, group delay, and reference place subsystems are:
:CALCulate{1-16}:PROCessing:ORDer Subsystem
:CALCulate{1-16}:REFerence Subsystem
:CALCulate{1-16}:PROCessing:ORDer Subsystem
:CALCulate{1-16}[:SELected]:GCOMpression Subsystem
:CALCulate{1-16}[:SELected]:GDELay Subsystem
:CALCulate{1-16}[:SELected]:TRANsform:TIME Subsystem
:SENSe{1-16}:CORRection:EXTension Subsystem
 
:CALCulate{1-16}[:SELected]:GCOMpression:INDicator[:STATe] <char1>
:CALCulate{1-16}[:SELected]:GCOMpression:INDicator[:STATe]?
Description
The command turns on/off the compression point indicator on the active trace of the indicated channel. The instrument must be in Power Sweep CW mode. The query outputs the on/off status of the compression point indicator on the active trace of the indicated channel. The instrument must be in Power-Sweep CW mode.
Cmd Parameters
<char1> 1 | 0 | ON | OFF
Query Parameters
<char1> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:GCOM:IND ON
:CALC1:GCOM:IND?
 
:CALCulate{1-16}[:SELected]:GCOMpression:PARameter <char1>
:CALCulate{1-16}[:SELected]:GCOMpression:PARameter?
Description
The command sets the S-Parameter for gain compression on the active trace of the indicated channel. The instrument must be in Power-Sweep Swept-Frequency mode. In the indicated channel, the query outputs the S-Parameter for gain compression on the active trace. The instrument must be in Power-Sweep Swept-Frequency mode. This value is not reset by an instrument reset.
Cmd Parameters
<char1> S11 | S12 | S21 | S22
Query Parameters
<char1> S11 | S12 | S21 | S22
Range
NA
Default Value
NA
Syntax Example
:CALC1:GCOM:PAR S22
:CALC1:GCOM:PAR?
 
:CALCulate{1-16}[:SELected]:GCOMpression:PVALue <NRf>
:CALCulate{1-16}[:SELected]:GCOMpression:PVALue?
Description
The command sets the gain compression point value on the active trace of the indicated channel. Instrument must be in Power Sweep CW mode or in Power Sweep Swept Frequency mode. The query outputs the gain compression point value on the active trace of the indicated channel. Instrument must be in Power Sweep CW mode or in Power Sweep Swept Frequency mode.
Cmd Parameters
<NRf> The input parameter is in dB.
Cmd Parameters
<NR3> The output parameter is in dB.
Range
MPNF
Default Value
1.000000E+000
Syntax Example
:CALC1:GCOM:PVAL 2.0E0
:CALC1:GCOM:PVAL?
 
:CALCulate{1-16}[:SELected]:GCOMpression:REFerence <char1>
:CALCulate{1-16}[:SELected]:GCOMpression:REFerence?
Description
The command sets the reference type for gain compression on the active trace of the indicated channel. Instrument must be in Power Sweep CW mode or in Power Sweep Swept Frequency mode. There are three types of gain compression references that can be set:
MAXGain = Maximum power gain at the highest point in the gain curve output.
PHOLd = The reference value will be that for the S-Parameter such as S21 set by the :CALCulate{1-16}[:SELected]:GCOMpression:PARameter command.
PINput = Input in dBm, this is the reference is value set for the selected S-Parameter which occurs when the input power is set to the value of gain compression reference value.
The query outputs the reference type for gain compression on the active trace of the indicated channel. Instrument must be in Power Sweep CW mode or in Power Sweep Swept Frequency mode.
Cmd Parameters
<char1> MAXGain | PHOLd | PINput
Query Parameters
<char1> MAXG | PHOL | PIN
Range
NA
Default Value
MAXG
Syntax Example
:CALC1:GCOM:REF PHOL
:CALC1:GCOM:REF?
 
:CALCulate{1-16}[:SELected]:GCOMpression:REFerence:VALue <NRf>
:CALCulate{1-16}[:SELected]:GCOMpression:REFerence:VALue?
Description
The command sets a reference level for gain compression on the active trace of the indicated channel. Instrument must be in Power Sweep CW mode or in Power Sweep Swept Frequency mode. The query outputs the reference level for gain compression on the active trace of the indicated channel. Instrument must be in Power Sweep CW mode or in Power Sweep Swept Frequency mode.
Cmd Parameters
<NRf> The input parameter is in dB.
Query Parameters
<NR3> The output parameter is in dB.
Range
MPNF
Default Value
0.000000E+000
Syntax Example
:CALC1:GCOM:REF:VAL -2.0E1
:CALC1:GCOM:REF:VAL?
 
:CALCulate{1-16}[:SELected]:GCOMpression:SNORmalize[:STATe] <char1>
:CALCulate{1-16}[:SELected]:GCOMpression:SNORmalize[:STATe]?
Description
The command toggles the self normalization mode on/off for the active trace of the indicated channel. The instrument must be in Power Sweep CW mode. The query outputs the on/off status of the self normalization mode on the active trace of the indicated channel. The instrument must be in Power Sweep CW mode.
Cmd Parameters
<char1> 1 | 0 | ON | OFF
Query Parameters
<char1> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:GCOM:SNOR 1
:CALC1:GCOM:SNOR?