VectorStar™ MS4640A SeriesMicrowave Vector Network Analyzer : SCPI Commands : :SENSe{1-16}:CORRection:COLLect:FLEXible Subsystem
 
:SENSe{1-16}:CORRection:COLLect:FLEXible Subsystem
The :SENSe{1-16}:CORRection:COLLect:FLEXible command defines the S-Parameter set for the existing calibration type on the given channel.
Calibration Option Subsystems
Related calibration option configuration and control subsystems are:
:CALCulate{1-16}:CORRection Subsystem - Adapters/Merge Calibration
:CALCulate{1-16}:EXTRaction Subsystem - Network Extraction
:CALCulate{1-16}:NXN Subsystem
:SENSe{1-16}:CORRection:COLLect:FLEXible Subsystem
:SENSe{1-16}:CORRection:COLLect:HYBRid Subsystem
 
:SENSe{1-16}:CORRection:COLLect:FLEXible:DEFine <char1>{,<char1>,...,<char1>}
:SENSe{1-16}:CORRection:COLLect:FLEXible:DEFine?
Description
The command defines the S-Parameter set for the existing calibration type on the given channel. At least one S-Parameter must be selected. Up to 16 S-Parameters may be selected in any sequence. The use of S-Parameters with Port 3 and/or Port 4 require a 4-Port VNA instrument. The query outputs the S-Parameter set defined for the existing calibration type on the given channel.
Cmd Parameters
<char1>{,<char2>,...<char16>} S11 | S12 | S21 | S22 | S13 | S31 | S23 | S32 | S33 | S14 | S41 | S24 | S42 | S34 | S43 | S44
Query Parameters
<char1>{,<char2>,...<char16>} S11 | S12 | S21 | S22 | S13 | S31 | S23 | S32 | S33 | S14 | S41 | S24 | S42 | S34 | S43 | S44
Range
NA
Default
S11
Syntax Example
:SENS1:CORR:COLL:FLEX:DEF S11, S12
:SENS1:CORR:COLL:FLEX:DEF?
S11,S12
 
:SENSe{1-16}:CORRection:COLLect:FLEXible[:STATe] <char1>
:SENSe{1-16}:CORRection:COLLect:FLEXible[:STATe]?
Description
The command toggles on/off the flexible definition of a calibration on the given channel. The query outputs the on/off status of the flexible definition of a calibration on the given channel.
Cmd Parameters
<char1> 1 | 0 | ON | OFF
Query Parameters
<char1> 1 | 0
Range
NA
Default
0
Syntax Example
:SENS1:CORR:COLL:FLEX ON
:SENS1:CORR:COLL:FLEX?