VectorStar™ MS4640A Series Microwave Vector Network Analyzers : Calibration Menus - 2-Port VNAs : AutoCal 1-Port Cal Setup - 2-Port VNAs
 
AutoCal 1-Port Cal Setup - 2-Port VNAs
 
AUTOCAL SETUP Menu - 1-Port Cal - 2-Port VNAs
Previous
 
AUTOCAL PORT Menu - 2-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | AutoCal | AUTOCAL PORT | 1-Port Cal | AUTOCAL SETUP
 
AUTOCAL SETUP Menu - One-Port Calibration - 2-Port VNAs
Modify Cal Setup
Select displays the MODIFY 1-PORT AUTOCAL SETUP dialog box. The dialog box provides control settings for 1-Port Calibration Ports and Port Left/Right Identification.
Port Selection (Read Only)
Displays the Port or Ports selected for the AutoCal procedure. The settings are determined in the MODIFY 1-PORT AUTOCAL SETUP dialog box.
Cal Type (Read Only)
Displays the Cal Type selected for the AutoCal procedure. The setting is defined in the AUTOCAL PORTS menu.
Port 1 Orientation (Read Only)
This read-only button only appears if Port 1 was selected in the MODIFY 1-PORT AUTOCAL SETUP dialog box. If available, shows the left/right assignment for Port 1.
Port 2 Orientation (Read Only)
This read-only button only appears if Port 2 was selected in the MODIFY 1-PORT AUTOCAL SETUP dialog box. If available, shows the left/right assignment for Port 2.
Begin Cal (AutoCal 1-Port Cal)
Starts the AutoCal procedure. On-screen dialogs and prompts provide user instructions for the selected calibration procedure. When the calibration is complete, the display returns to the CALIBRATION [TR] menu when the Cal Status button is enabled and set to ON.
 
MODIFY 1-PORT AUTOCAL SETUP Dialog Box - 2-Port VNAs
 
Previous
AUTOCAL SETUP Menu - 1-Port Cal - 2-Port VNAs
Navigation
MAIN | Calibration | CALIBRATION | Calibrate | CALIBRATE | AutoCal | AUTOCAL PORT | 1-Port Cal | AUTOCAL SETUP | Modify Cal Setup | MODIFY 1-PORT AUTOCAL SETUP Dialog Box
 
MODIFY 1-PORT AUTOCAL SETUP Dialog Box - 2-Port VNAs
1. AutoSense not available. Port 1 and Port 2 selected for calibration. Port 1 = Left and Port 2 = Right.
2. Only Port 1 selected for calibration. Port 1 = Left.
3. Only Port 1 selected for calibration. Port 2 = Left.
 
Auto Sense Cal Port(s)
The Auto Sense selection check box is not available in AutoCal 1-Port.
1-Port Cal Port(s)
Select either one or both of the available ports.
Port 1 Only
See Callout #1 in Figure: MODIFY 1-PORT AUTOCAL SETUP Dialog Box - 2-Port VNAs.
If selected, the Port 1 Cal Left/Right area is available.
Port 1 and Port 2
See Callout #2 in Figure: MODIFY 1-PORT AUTOCAL SETUP Dialog Box - 2-Port VNAs.
If selected, the Port 1 Cal Left/Right area is available.
Port 2 Only
See Callout #3 in Figure: MODIFY 1-PORT AUTOCAL SETUP Dialog Box - 2-Port VNAs
If selected, the Port 2 Cal Left/Right area is available.
Port 1 Cal Left/Right Radio Buttons
Port 1 above must be selected to make the radio buttons available. If available, allows the following port left/right options:
Left = Port 1
Right = Port 1
See Callout #2 in Figure: MODIFY 1-PORT AUTOCAL SETUP Dialog Box - 2-Port VNAs.
Port 2 Cal Left/Right Radio Buttons
Port 2 above must be selected to make the radio buttons available. If available, allows the following port left/right options:
Left = Port 2
Right = Port 2
See Callout #3 in Figure: MODIFY 1-PORT AUTOCAL SETUP Dialog Box - 2-Port VNAs.