Note | The appearance and button availability of the calibration menus depends on the settings established in the CAL SETUP, CAL METHOD, LINE TYPE menus and in the associated dialog boxes that appear from the Edit Cal Params button. |
Modify Cal Setup Select displays the CAL SETUP menu where the Edit Cal Params button provides access to the TRANSMISSION FREQUENCY RESPONSE CAL SETUP dialog box for the selected calibration method and line type. Example transmission frequency response calibration dialog boxes are available below: A summary table of all transmission frequency response calibration configuration dialog boxes is available here: | |
Port Selected Read-only display of the ports selected for the pending calibration. Completion Menu Buttons For this example menu, the Thru/Recip and Isolation (Optional) buttons link to completion submenus where additional calibration procedures are performed. For example, the Thru/Recip button (shown below at #1) links to the THRU/RECIP submenu. As each procedure is completed, the submenu button is marked with a completion checkmark. When all the procedures on the submenu are completed, use the Back button to return to the TRANS. RESPONSE menu. The Thru/Recip button (shown above at #2) is now marked with a completion checkmark. Thru/Recip When selected, displays the THRU/RECIP menu. When all tasks are complete, return to the TRANS. RESPONSE menu. Isolation (Optional) When selected, displays the ISOLATION menu. When all tasks are completed, return to the TRANS. RESPONSE menu. Done This button is unavailable until all calibration tasks have been successfully completed. When available, select the button to return to the CALIBRATION menu when the Cal Status is set to ON. Abort Cal Select aborts the current calibration and returns to the CALIBRATION menu. |
TRANSMISSION FREQUENCY RESPONSE CALIBRATION SETUP Dialog Box |
Cal Method Line Type | Dialog Box Input Selections and Controls |
SOLT/R Coaxial | See full description above at TRANS. FREQ. RESP. CAL. SETUP (SOLT/R - Coaxial) Dialog Box. Select Direction and Ports: Port 1, Port 2, Port 1 and Port 2 Reference Impedance (Ohms) |
SOLT/R Non-Dispersive | Same controls and functions as SOLT/R Coaxial. |
SOLT/R Waveguide | Same controls and functions as SOLT/R Coaxial with the following changes: Waveguide Kit: User-Defined1 to User-Defined32 Waveguide Info button: Displays USER DEFINED WAVEGUIDE dialog box. |
SOLT/R Microstrip | Same controls and functions as SOLT/R Coaxial with the following changes: Microstrip Kit: 10 Mil Kit, 15 Mil Kit, 25 Mil Kit, User-Defined1 to User-Defined32 Microstrip Info button: Displays info dialog box for selected calibration method and kit. • Typical MICROSTRIP INFO Dialog Box - 2-Port VNAs |
SSLT Coaxial | Same controls and functions as SOLT/R Coaxial above. |
SSLT Non-Dispersive | Same controls and functions as SOLT/R Coaxial above. |
SSLT Waveguide | Same controls and functions as SOLT/R Coaxial with the following changes: Waveguide Kit: User-Defined1 to User-Defined32 Waveguide Info button: Displays USER DEFINED WAVEGUIDE dialog box. • Typical STANDARD INFO Dialog Box - 2-Port VNAs |
SSLT Microstrip | Same controls and functions as SOLT/R Coaxial with the following changes: Microstrip Kit: 10 Mil Kit, 15 Mil Kit, 25 Mil Kit, User-Defined1 to User-Defined32 Microstrip Info button: Displays info dialog box for selected calibration method and kit. • Typical MICROSTRIP INFO Dialog Box - 2-Port VNAs |
SSST Coaxial | Same controls and functions as SOLT/R Coaxial above. |
SSST Non-Dispersive | Same controls and functions as SOLT/R Coaxial above. |
SSST Waveguide | Same controls and functions as SOLT/R Coaxial with the following changes: Waveguide Kit: User-Defined1 to User-Defined32 Waveguide Info button: Displays USER DEFINED WAVEGUIDE dialog box. |
SSST Microstrip | Same controls and functions as SOLT/R Coaxial with the following changes: Microstrip Kit: 10 Mil Kit, 15 Mil Kit, 25 Mil Kit, User-Defined1 to User-Defined32 Microstrip Info button: Displays info dialog box for selected calibration method and kit. • Typical MICROSTRIP INFO Dialog Box - 2-Port VNAs |
LRL/LRM Coaxial | The LRL/LRM calibration method is not available for transmission frequency response calibrations. |
LRL/LRM Non-Dispersive | |
LRL/LRM Waveguide | |
LRL/LRM Microstrip |