Cable and Antenna Analyzer for Anritsu’s RF and Microwave Handheld Instruments : Cable and Antenna Analyzer : Using Trace Math
  
Using Trace Math
Trace Math menu is an excellent tool for comparing two traces. It is possible to upload traces using Line Sweep Tools, store those traces in memory, and compare the stored trace with a more recent trace. It is also possible to use the trace math features to normalize the trace and obtain a reference for S21 measurements where a full 2-port calibration is not needed. The trace math menu can be used to add or subtract logarithmic data (multiply and divide linear data).
The Trace Math menu is accessed by pressing the Shift key and the Trace (5) key.
Example 1
Comparing the Return Loss data for two five meter cables, Cable A and Cable B (Figure: Trace Menu Showing Trace – Memory).
1. Press the Measurement main menu key and press the Return Loss submenu key.
2. Connect Cable B to the BTS Master RF Out port.
3. Press the Shift key and the Trace (5) key.
4. Press the Copy Trace To Display Memory submenu key.
5. Connect Cable A to the BTS Master RF Out port and perform the same Return Loss measurement over the same frequency range.
6. Turn Trace Overlay submenu key On to view the Return Loss measurement of Trace B.
7. Press the Trace + Memory submenu key to view the Return Loss measurement of Trace A and Trace B at the same time.
8. Press the Trace - Memory submenu key to look at the difference in Return Loss between the two cables.
Trace Menu Showing Trace – Memory
Example 2
Using the Trace Menu to normalize a trace for Transmission measurement. Please note that normalizing the trace will not give the same accuracy as a calibrated 2-port measurement.
Note 
The instrument incorporates internal gain stages that can negatively affect measurement accuracy if used uncalibrated. This effect can be exaggerated when using trace math to compare measurements with different gains.
1. Connect a through cable between Ports 1 and 2 for the normalization procedure.
1. Press the Measurement main menu key. Press the 2-Port Gain submenu key.
2. Under the Freq/Dist main menu key, use the Start Freq and Stop Freq submenu keys to set the Start and Stop frequencies.
3. Press the Shift key and the Trace (5) key.
4. Press the Copy Trace To Display Memory submenu key.
5. Press the Trace - Memory submenu key to see the trace normalized around 0 dB (Figure: Normalizing the Trace Using Trace Math).
Normalizing the Trace Using Trace Math