Spectrum Analyzerfor Anritsu RF and Microwave Handheld Instruments : Interference Analyzer (Option 25) : Measurements Menu : Emission Mask Menu
  
Emission Mask Menu
Key Sequence: Measurements > Spectrum > Masks and C/I > Emission Mask
IA Emission Mask Menu
This submenu controls the setup and display of the emission mask. The emission mask is an upper segmented limit line. It must have more than two nodes, which mean at least two segments. The spectrum emission mask measurement includes the in-band and out-of-band emissions.
 
Emission Mask
This submenu controls the setup and display of the emission mask. The spectrum emission mask measurement includes the in‑band and out‑of‑band emissions.
Emission Mask
On Off:
Turns On/Off the Emission Mask graph and table display.
Recall Limit as Emission Mask
Opens a folder of limit lines to select a limit line for use as the Emission Mask.
Ref Power
Peak Channel
Press to display the Reference Power as Peak or Channel. When Channel is selected, the Reference Power value is the integral of the individual peaks within the channel.
Channel Width
Channel Width is set within the Signal Standard. Use this button to adjust the width as desired.
 
 
Peak Marker On/Off
Turning on this feature displays a peak marker within an Emission Mask segment. For example, if the Emission Mask had seven segments, then seven peak markers would be displayed. Passing peak markers are light blue in color, while peak markers that exceed the mask limit are red in color.
Back
Returns to the Masks and C/I Menu.