Cable and Antenna Analyzer Measurement Guide : Cable and Antenna Analyzer : Marker Menu
 
Marker Menu
Key Sequence: Marker
Marker Menu
Marker
Press the marker key the select M1 to M6 on the touch screen to choose the active marker. The underlined marker number is the active marker.
On/Off
Turns the selected marker underlined in the Marker key On or Off.
Delta
Turns on a delta marker.
Marker to Peak
This submenu key places the currently active marker on the highest signal amplitude currently displayed on screen.
Marker to Valley
This soft key places the currently active marker on the lowest signal amplitude currently displayed on screen.
Peak/Valley Auto (Markers 1 through 4 Active)
Automatically set Marker 1 to peak and Marker 2 to valley. Note: Not available in DTF measurements.
Marker Options (Marker 5 or 6 Active)
Displays the Marker Options submenu.
Marker 5 Active
Peak Between M1&M2: Place Marker 5 at the peak between Marker 1 and Marker 2.
Valley Between M1&M2: Place Marker 5 at the valley between Marker 1 and Marker 2.
Marker Unbounded Bounded:
Bound Marker 5 with Marker 1 and Marker 2. Used in conjunction with Bounded Limit Lines (see Limit Menu).
Marker 6 Active
Peak Between M3&M4:
Place Marker 6 at the peak between Marker 3 and Marker 4.
Valley Between M3&M4: Place Marker 6 at the valley between Marker 3 and Marker 4.
Marker Unbounded Bounded: Bound Marker 6 with Marker 3 and Marker 4. Used in conjunction with Bounded Limit Lines (see Limit Menu).
Marker Table
Causes a table to be displayed below the sweep window. The table is automatically sized to display all markers that are turned on. In addition to the marker frequency and amplitude, the table also shows delta frequencies and amplitude deltas for all markers that have deltas entered for them.
All Markers Off
Turns off all markers.