Cell Master User Guide : File Management : File Menu : Save On... Menu
 
Save On... Menu
Key Sequence: Shift > File (7) > Save On Event
Save On Menu
In Spectrum Analyzer mode, this menu is used to automatically save measurements to the current save location after any of the following events:
...Crossing Limit
On Off
When this submenu key is set to On, measurement data is automatically saved when the trace crosses a limit line that has been specified with the Limit menu.
...Sweep Complete
On Off
When this key is set to On, measurement data is automatically saved after the current sweep is completed.
Save Then Stop
On Off
Set this key to On to stop the sweep after a measurement is saved. If this setting is Off and Sweep Complete is On, a measurement is saved after every sweep.
Clear All
Press this key to turn Off all three Save On Event keys:
Crossing Limit
Sweep Complete
Save Then Stop
Back
Press this key to return to the File Menu.