These Tracking Generator options include a CW Generator. The CW Signal Generator provides a continuous wave (CW) signal from the GENERATOR/RF OUT port of the instrument.
The following example displays the generator CW signal on the instrument display.
2. Set the Mode to Spectrum Analyzer. Refer to General Information for details.
3. Preset the instrument (Shift+1 > Preset).
4. Set the spectrum analyzer frequency range using the Frequency (Freq) menu and submenus. Refer to the Spectrum Analyzer Measurement Guide (10580‑00349) for additional information.
5. Press the Shift key, then the Measure (4) key. Press the Generator key.
6. Press the Output Power button and set the output power (at the output connector before any external attenuation).
7. Set the Generator Mode to CW and set the CW Frequency.
Note
Confirm that the CW Frequency is set within the spectrum analyzer span, or the signal will not be displayed on the instrument.
8. Turn Generator OutputOn. The output power turns on at the level that has been set previously with the Output Power key.