Spectrum Analyzer : Sweep Menu : Gate Setup Menu (Option 90)

Gate Setup Menu (Option 90)
Key Sequence: Shift > Sweep (3) key > Gated Sweep Setup
Gate Source is currently only available using an External trigger signal that can be input using the instrument’s Ext Trigger In connector.
Sets the start of the gated sweep indicated by the left border of the blue dashed rectangle shown in the bottom graph of Figure: “Gated Sweep Example”.
Sets the length of the gate and is reflected on the zero span graph by the width of the blue rectangle as shown in Figure: “Gated Sweep Example”.
Opens the Gate View Setting submenu. Allows a user to independently change the RBW, VBW and sweep time of the zero span or gate view (bottom graph).
Zero Span RBW: Sets the resolution bandwidth of the zero span graph.
Zero Span VBW: Sets the video bandwidth of the zero span graph.
Zero Span Time: Sets the sweep time of the zero span graph.
Back: Returns to the Gate Setup menu.
Returns to the “Sweep Menu” and also changes the Gated Sweep Setup view back to the full screen Spectrum view. The Gated Sweep settings are retained and applied to the spectrum.