LMR Master™S412E User Guide : Error Messages : Reset Options
 
Reset Options
You can reset your LMR Master to Factory Defaults or use a Master Reset to return to the FULL Factory Default condition from the instrument menu or from the Off condition.
Reset Via Instrument Menus
To reset the instrument from the menu:
1. On the LMR Master menu system, press the Shift key, then the System (8) key to open the System menu.
2. Press the System Options submenu key to open the System Options menu.
3. Press the Reset submenu key to open the Reset menu (refer to Reset Menu).
4. From the Reset menu, press either the Factory Defaults submenu key or the Master Reset submenu key.
Caution
Using Master Reset erases all of the user-saved setups and measurement traces and returns the LMR Master to a full Factory Default condition.
Reset from OFF Condition
You can also reset the LMR Master by turning it Off and then restarting under one of the following conditions:
Factory Defaults Reset:
To preserve user files and reset to factory defaults:
1. Hold the Esc button while pressing and releasing the On/Off button.
2. Continue holding the Esc button until the Anritsu splash screen appears.
3. Release the Esc button. The LMR Master starts up with many Factory Default settings (refer to Factory Defaults). Throughout this appendix, this sequence is abbreviated as Factory Defaults (Esc + On).
Master Reset:
Caution
Using Master Reset, erases all of the user-saved setups and measurement traces and returns the LMR Master to a full Factory Default condition.
To force a master resetting of the instrument:
1. Hold the 8 key in the number keypad (also referred to as the System (8) key) while pressing and releasing the On/Off button.
2. Continue holding the 8 key until the Anritsu splash screen appears.
3. Release the key. The LMR Master starts up in FULL Factory Default condition (refer to Master Reset). Throughout this appendix, this sequence is abbreviated as Master Reset (System + On).