SCPI CommandsPlease refer to your instrument’s programming manual for the use of SCPI commands. This set of commands prepares the PIM instrument hardware for the selected measurement. They activate the specified measurement and sets the instrument to a wait‑for‑sweep mode, waiting for an :INITiate command to take a measurement.A typical command set for the PIM Master would include:[SENSe]:PIManalyzer:FREQuency:F2 1990000000 (Sets F2 to 1990 MHz)
[SENSe]:PIManalyzer:AUTorange 1 (Sets Amplitude to Auto Range)[SENSe]:PIManalyzer:IMD:ORDer 3 (Sets center frequency of Rx to IM3)[SENSe]:PIManalyzer:OUTPut:POWer 2 (Sets power to 20 Watts)[SENSe]:PIManalyzer:TEST:DURation 10 (Sets the power on time)INITiate:PIManalyzer:MEASure PIM (Starts PIM measurement)