Bias Tee (On/Off) Press this soft key to open the Bias Tee Menu to turn On or Off the Bias Tee voltage. Int voltage P2 Press this soft key to set the internal bias tee voltage that is directed onto the center conductor of Port 2. The available range is from 12.0 V to 32.0 V in increments of 0.1 . Use the arrow keys, the rotary knob, or the number keypad to change the setting. When using the number keypad, press the soft key for voltage units (V), or press the Enter key. Press the Esc key to exit without changing the setting. Int Current Limit P2 Press this soft key to set the internal bias tee current limit for the voltage that is set at Port 2. The available range is from 0 mA to 450 mA in steps of 1 mA. This current limit sets the trip point for the bias tee for this port. When using the number keypad, the soft key menu displays 2 choices for units: A or mA. Use the arrow keys, the rotary knob, or the number keypad to change the setting. When using the number keypad, press the soft key for current units (A or mA), or press the Enter key to use mA. Press the Esc key to exit without changing the setting. Back Press this soft key to the Sweep menu. |
Off Press this soft key to turn Off the Bias Tee function. On Press this soft key to turn ON the Bias Tee function. Back Press this soft key to return to the Bias Tee Setup Menu without changing the current Bias Tee setting. |