VNA Masterâ„¢Programming Manual : Power Monitor Commands : :FETCh:PMONitor Subsystem
 
:FETCh:PMONitor Subsystem
This subsystem contains commands to fetch the power monitor reference power level data, zero power level data and the displayed power level data.
:FETCh:PMONitor Subsystem
Keyword
Parameter Data or Units
:FETCh
 
:PMONitor
 
:RELative
:ZERO
Displayed Data
:FETCh:PMONitor:DATA?
Title
Displayed Data
Description
Fetches the displayed power level data. The returned value is returned in the units that have been set with the Units command (:CALCulate:PMONitor:UNITs DBM|WATT|DB|PERCent).
If units are in dBm, then the returned value is in dBm.
If units are in percent, then the returned value is in percent.
If units are in dB, then the returned value is in dB.
If units are in watts, then the returned value is in 0.1 uW (0.1 microwatt) increments. For example:
Returned Value
Measurement Value
1
0.1 microwatt
10
microwatt
25000
2500 microwatts or 2.50 milliwatts
12000000
1200000 microwatts or 1.2000000 watts
Syntax
:FETCh:PMONitor:DATA?
Cmd Parameter
NA (query only)
Query Response
<NR3> (depends on set unit)
Example
To fetch the displayed power level data:
:FETCh:PMONitor:DATA?
Front Panel Access