VNA Master™ User Guide Model MS20xxC Vector Network Analyzer with Spectrum Analyzer : Quick Start Guide : Memory Profile and Security Issues
 
Memory Profile and Security Issues
This section describes the profiles of the various types of memory that are used in the MS20xxC VNA Master and the associated security issues that are related to those memory devices.
The MS20xxC has 1 GB of Flash non‑volatile memory, has EEPROM memory, and has sufficient DRAM volatile memory for normal operation. The instrument is supplied with a USB memory device that plugs into the USB Type A connector. The MS20xxC does not have a hard drive or any other type of volatile or non‑volatile memory.
The following sections describe how memory is used in the VNA Master and how it can be erased.
Internal Flash Memory
This memory space is used to store the instrument firmware and factory calibration, and can be used to store measurements and setups that are saved by the user.
Saved measurements and setups stored in the Flash memory are all deleted by the master reset process that is described in MS20xxC Master Reset Instructions.
Note 
With Option 7 enabled, the user is not able to store any measurements or setups to this internal flash memory.
USB Memory Device
A USB memory device is not required for operating the instrument. You can store measurement and setup data directly on a USB memory device, or you may transfer the contents of internal memory to the USB device.
Note 
With Option 7 enabled, the user does not have access to the instrument’s internal memory and can only save to or load from a USB device.
RAM Memory
This is volatile memory that is used to store many parameters needed for the normal operation of the MS20xxC along with current measurements. This memory is reset whenever the instrument is restarted.
EEPROM
This memory holds information such as the model number, serial number, and calibration data for the instrument. Also stored here are the operating parameters, such as frequency range, that are set by the user. During the master reset process, all operating parameters that are stored in the EEPROM are set to standard factory default values.
MS20xxC Master Reset Instructions
1. Turn on the instrument.
2. Press the Shift key, then the System (8) key.
3. Press the System Options key.
4. Press the Reset key.
5. Press the Master Reset key.
6. A dialog box is displayed on the instrument screen to warn that all settings will be returned to factory default values, and that all user files will be deleted.
7. Press the Enter key to execute the master reset, or press the Esc key to abort.
8. After several seconds (which can grow to several minutes if a very large number of measurements have been saved in the instrument), the instrument reboots.  
Caution
Use of Master Reset (System + On), will permanently erase all user saved setups and measurement traces and will return the VNA Master to a full Factory Default condition. Be sure to make backup copies of any important files before proceeding with a Master Reset.