Spectrum Analyzer Measurement Guide : Bias Tee (Option 10)
 
Bias Tee (Option 10)
 
Introduction
Bias Tee Menu
Note 
Not all instrument models offer every option. Please refer to your instrument Technical Data Sheet for available options.
Introduction
Option 10 provides a bias tee that is installed inside the instrument. Refer to the Vector Network Analyzer Measurement Guide (10580-00289) for complete description of this option. The bias arm is connected to a 12 VDC to 32 VDC power source that can be turned on as needed to place the voltage on the center conductor of the instrument’s RF In port. This supply of bias implies it is mostly useful when conducting two-port transmission measurements. This voltage can be used to provide power to block down-converters in satellite receivers and can also be used to power some tower-mounted amplifiers.
The bias can be turned on only when the instrument is in transmission measurement, return loss, cable loss, VSWR, DTF, Spectrum Analyzer, Channel Scanner, or Interference Analyzer mode. When bias is turned on, the bias voltage and current are displayed in the lower left corner of the display. The 12 VDC to 32 VDC power supply is designed to continuously deliver a maximum of 6 Watts. Refer to Figure: Variable Bias Tee for a sample bias tee connection.
­
Variable Bias Tee