VNA Masterâ„¢Programming Manual : VNA Commands : :CALCulate:SMOothing Subsystem
 
:CALCulate:SMOothing Subsystem
This subsystem contains functions for trace smoothing.
Smoothing
:CALCulate<Tr>:SMOothing:APERture
Description
Sets the smoothing percentage for the given trace <Tr>. The query form of the command returns the current smoothing percentage. <Tr> is the trace number in the range 1 to 4. If no trace number is specified, then the <Tr> parameter defaults to trace number 1. Note that setting the smoothing also sets the given trace as the active trace if it is not already active.
Syntax
:CALCulate<Tr>:SMOothing:APERture <integer>
:CALCulate<Tr>:SMOothing:APERture?
Cmd Parameter
<NR1> <integer>
Query Response
<NR1> <integer>
Default Value
0
Range
0 to 20
Front Panel Access
Shift 4 (Measure), Smoothing %