VNA Master™Programming Manual : VNA Commands : :SYSTem Subsystem
 
:SYSTem Subsystem
The commands in this subsystem relate to the current operating state of the instrument.
 
System Alternate State
:SYSTem:ALTernate:STATe
Description
Sets the state of the Measurement View to be either VNA or Field. The query form of this command returns a 0 for VNA and a 1 for Field.
Syntax
:SYSTem:ALTernate:STATe VNA|FIELD
:SYSTem:ALTernate:STATe?
Cmd Parameter
<char> VNA|FIELD
Query Response
char 0|1
Default Value
VNA
Example
To set the Measurement View to Field:
:SYSTem:ALTernate:STATe FIELD
To set the Measurement View to VNA:
:SYSTem:ALTernate:STATe VNA
Front Panel Access
Shift-8 (System), Application Option, Meas Menu
Motherboard Temperature
:SYSTem:MBTemperature?
Description
This command returns the current mother board temperature in degrees Celsius.
Syntax
:SYSTem:MBTemperature?
Cmd Parameter
NA (query only)
Query Response
<NR3> (degree Celsius)
Front Panel Access
NA
 
Preset
:SYSTem:PRESet
Description
This command restores all application parameters to their factory preset values. This command does not modify system parameters such as Ethernet configuration, language, volume, or brightness.
Note: After issuing this command, wait 30 seconds for the instrument to complete the preset condition before sending another command.
Syntax
:SYSTem:PRESet
Cmd Parameter
NA
Query Response
NA (no query)
Related Command
*RST
Front Panel Access
Shift‑1 (Preset), Preset