VNA Master™Programming Manual : Vector Voltmeter Commands : :MMEMory:LOAD Subsystem
 
:MMEMory:LOAD Subsystem
The Mass MEMory LOAD subsystem contains commands to transfer from the mass memory device to the internal memory.
Recall Setup
:MMEMory:LOAD:STATe
Description
Recalls a previously stored setup from the current storage location. The saved setup that is to be loaded is specified by <filename>. <filename> must be enclosed in either single quotes (‘ ’) or double quotes (“ ”) and must include the extension “.stp”. The <integer> parameter is not currently used, but it must be sent. Send a value of 1.
Note 
When recalling setups for modes other than the current mode, Anritsu recommends that you first change modes, then recall the setup.
Use :INSTrument:NSELect or :INSTrument[:SELect] to change the instrument state, then use :MMEMory:LOAD:STATe to recall the setup.
Syntax
:MMEMory:LOAD:STATe <integer>, <filename>
Cmd Parameter
<integer>, <string> (1, file name)
Query Response
NA (no query)
Related Command
:MMEMory:STORe:STATe
:MMEMory:MSIS INTernal|USB
Front Panel Access
Shift‑7 (File), Recall
 
Recall Measurement
:MMEMory:LOAD:TRACe
Description
Recalls a previously stored measurement trace from the current storage location. The saved measurement trace that is to be loaded is specified by <filename>. <filename> must be enclosed in either single quotes (‘ ’) or double quotes (“ ”) and must contain a file extension of “.mna”. Note that the trace that is specified by <filename> must be available at the current mass storage location. The <integer> parameter is not currently in use, but it must be sent. Send a 1. File Extensions: “.mna”.
Syntax
:MMEMory:LOAD:TRACe <integer>, <filename>
Cmd Parameter
<integer>, <string> (1, file name)
Query Response
NA (no query)
Example
To recall trace with file name “trace”:
:MMEMory:LOAD:TRACe 1,“trace.mna”
Related Command
:MMEMory:STORe:TRACe
:MMEMory:MSIS INTernal|USB
Front Panel Access
Shift‑7 (File), Recall