PIM Master™ Passive Intermodulation Analyzer : Swept PIM : Setup Menu
  
Setup Menu
Key Sequence: Setup
Setup Menu for Swept PIM
High (37–46 dBm) Output Power
Press this submenu key to set the output power level within the higher range for both carrier frequencies, F1 and F2.
Low (25–37 dBm) Output Power
Press this submenu key to set the output power level within the lower range for both carrier frequencies, F1 and F2.
Note: Power output value from the test port is twice the set output value. For example, when Setup > Output Power is set to 20 W, total output power is 40 W.
20 W (F1) + 20 W (F2) = 40 W
Test Duration
Press this submenu key to set the length of time that the test port will deliver power after Measurements > Test is pressed. Range for the test is 1 second to 1200 seconds. Use the Arrow keys or the number keypad. If you use the number keypad, then the Units menu of the Test Duration menu is displayed.
 
Measurements Menu
Refer to Measurements Menu.
Marker Menu
The Marker menu is not available for Swept PIM measurements. Although the Marker menu is not available, Marker 1 is always On and always set to the trace peak. Refer to Marker 1.