High (37–46 dBm) Output Power Press this submenu key to set the output power level within the higher range for both carrier frequencies, F1 and F2. Low (25–37 dBm) Output Power Press this submenu key to set the output power level within the lower range for both carrier frequencies, F1 and F2. Note: Power output value from the test port is twice the set output value. For example, when Setup > Output Power is set to 20 W, total output power is 40 W. 20 W (F1) + 20 W (F2) = 40 W Test Duration Press this submenu key to set the length of time that the test port will deliver power after Measurements > Test is pressed. Range for the test is 1 second to 1200 seconds. Use the Arrow keys or the number keypad. If you use the number keypad, then the Units menu of the Test Duration menu is displayed. |