Torque values are listed below for connectors used with the 365xx-x mechanical calibration kits. If you are not sure which connector types are available on your instrument, consult the instrument Technical Data Sheet available on the web at www.anritsu.com.
Connector Torque Settings and Recommended Tools 
Connector
Type/Size
Wrench Size
Torque Setting

Recommended Tools
3.5 mm1 2
8 mm
5/16 in
0.9 N·m
8 lbf·in
01-201 5/16 in Torque End Wrench
01-204 5/16 in or 8 mm Open End Wrench
GPC-7
19 mm
3/4 in
12 lbf·in 1.36 N·m
01-200 3/4 in Torque End Wrench
K
(2.92 mm)
8 mm
5/16 in
0.9 N·m
8 lbf·in
01-201 5/16 in Torque End Wrench
01-204 5/16 in or 8 mm Open End Wrench
V
(1.85 mm)
8 mm
5/16 in
0.9 N·m
8 lbf·in
01-201 5/16 in Torque End Wrench
01-204 5/16 in or 8 mm Open End Wrench
K, V, 2.4 mm
(Test Port Adapter)
1/2 in
4.1 N·m
36 lbf·in
1/2 in Torque End Wrench
1/2 in Open End Wrench
W
(1 mm)
6 mm
0.45 N·m
4 lbf·in
01-504 6 mm Torque End Wrench and 7 mm Open End Wrench
0.8 mm
6 mm
0.45 N·m
4 lbf·in
01-504 6 mm Torque End Wrench and 7 mm Open End Wrench

1 3.5 mm, 2.92 mm and K connectors are electrically compatible with 3.5 mm and have the same connector nut size (8 mm) and torque requirement. Refer to Table: RF Connector Characteristics for connector compatibility information.

2 Although the 2.4 mm, 1.85 mm, and V connectors have the same connector nut size (8 mm) and torque requirement, they are not mechanically compatible with 3.5 mm. Refer to Table: RF Connector Characteristics for connector compatibility information.