VectorStarâ„¢ MS464xB Series Microwave Vector Network Analyzer Calibration and Measurement Guide : AutoCal Calibration : Available AutoCal Modules and Adapters
 
Available AutoCal Modules and Adapters
The Precision and Standard AutoCal modules are show in the figure below.
AutoCal Automatic Calibrator Calibration Modules
1. 36581KKF Standard AutoCal Module (at left).
2. 36585-Series Precision AutoCal Module (at right).
Anritsu has both Precision AutoCal modules as well as a Standard AutoCal module. The table below shows the available modules.
Precision and Standard AutoCal Kits and Adapters 
Part Number
Name
Frequency Range and Connectors
Precision AutoCal Kits
36585V-Series Precision AutoCal Modules can be used with MS4642B or MS4644B VNAs with adapters. The calibration range is limited to that of the VNA.
Only the 36585V-2MF and 36585K-2MF Precision AutoCal Modules can be used with the 366xX-1 Verification Kits.
The 36585K-Series Precision AutoCal Kits can be used with MS4645B or MS4647B VNAs with adapters. The calibration range is limited to 40 GHz.
36585K-2M
Precision AutoCal Kit, K (m-m), 2-port
70 kHz to 40 GHz, K (m) to K (m)
36585K-2F
Precision AutoCal Kit, K (f-f), 2-port
70 kHz to 40 GHz, K (f) to K (f)
36585K-2MF
Precision AutoCal Kit, K (m-f), 2-port
70 kHz to 40 GHz, K (m) to K (f)
36585V-2M
Precision AutoCal Kit V (m-m), 2-port
70 kHz to 70 GHz, V (m) to V (m)
36585V-2F
Precision AutoCal Kit V (f-f), 2-port
70 kHz to 70 GHz, V (f) to V (f)
36585V-2MF
Precision AutoCal Kit V (m-f), 2-port
70 kHz to 70 GHz, V (m) to V (f)
Standard AutoCal Kit
The 36581KKF Standard AutoCal Module can be used with any MS464xB VNA with adapters as required on the Test Ports and/or Module.
The calibration range is limited to 20 GHz.
36581KKF
Standard AutoCal Kit
40 MHz to 20 GHz, K (m) to K (f)
Standard AutoCal Kit Matched Adapters
Matched adapters are recommended only for use with the 36581KKF Standard AutoCal Module. The standard AutoCal module can be ordered with connector adapters which assume the use of K (m) test port cables. These adapters can be also used on the Precision AutoCal modules, but with some degradation of precision. See Calibration and Measurement Overview for available calibration modules, tools, and accessories.
36583K
Matched K Adapters Set, 2 adapters
40 MHz to 20 GHz, K (f) to K (m) adapters
36583L
Matched 3.5 mm Adapters Set, 4 adapters
40 MHz to 20 GHz,
K(f) to 3.5 mm (m) adapters, 2 each
K (f) to 3.5 mm (f) adapters, 2 each
36583S
Matched SMA Adapters Set, 3 adapters
40 MHz to 20 GHz,
K (f) to SMA (m) adapter, 1 each
K (f) to SMA (f) adapters, 2 each