VectorStar™ MS464xB Series Microwave Vector Network Analyzer Calibration and Measurement Guide : PulseView™ (Option 35 and Option 42) : Point-in-Pulse Measurements
 
Point-in-Pulse Measurements
This section describes point-in-pulse measurement examples in more detail and provides guidance on how to select parameters of importance. Point-in-pulse measurements are often used when some portion of the pulse is not of interest and only the averaged result over a specified measurement window is of concern.
Pulsing is often done for thermal reasons or when some measure of basic RF performance in some non-dynamic portion of the pulse is the main measurement requirement, or for both of these reasons. Typically, this measurement is done over swept power or frequency.
Multiple point-in-pulse measurements may be run on different VNA channels where the frequency (or power) sweep parameters and pulse parameters are the same, but the point in the pulse where data is taken is different. In this way, information regarding the sweep parameter and the intra-pulse behavior can be acquired nearly simultaneously.
A pair of power sweeps (perhaps set up in two VNA channels) where the effect of pulse width will be examined provides our example. This example neglects transient effects near the beginning and end of the pulse, but will reveal whether there are any effects related to an increase in pulse width (and, in this case, duty cycle).
In this example, only RF is being pulsed, and the PRI is 0.1 ms. Pulse widths of 20 μs and 80 μs are used, and the point-in-pulse measurement is set to roughly the middle of each pulse with a width of 10 μs. The amount of time between initiation of the stimulus pulse and the measurement is different in the two cases, so the effect may be a convolution of actual measurement time and the duty cycle presented to the DUT. In this simple example, only one pulse generator is used. A delay relative to T0 is used for clarity.
Sweep 1:
Details of Sweep 1 Timing Relationships in the Point-in-Pulse Setup Example
PRI: 100 μs
Pulse generator 4: delay 50 μs, width 20 μs
Point-in-pulse position: 55 μs
Point-in-pulse measurement width: 10 μs
Power sweep: –10 to 10 dBm, 41 points, frequency CW 18 GHz
Sweep 2:
Details of Sweep 2 Timing Relationships in the Point-in-Pulse Setup Example
PRI: 100 μs
Pulse generator 4: delay 10 μs, width 80 μs
Point-in-pulse position: 55 μs
Point-in-pulse measurement width: 10 μs
Power sweep: –10 to 10 dBm, 41 points, frequency CW 18 GHz
Approximate timing relationships for this measurement are shown in Figure: Details of Sweep 1 Timing Relationships in the Point-in-Pulse Setup Example and Figure: Details of Sweep 2 Timing Relationships in the Point-in-Pulse Setup Example. DUT (amplifier) output power is the measurement of interest, so the pulse measurement will be processed only on b2.
The dialog boxes and settings for these examples are shown in Figure: Sweep 1 Settings in Pulse Configuration Dialog Box, Figure: Sweep 2 Settings in the Pulse Configuration Dialog Box, and Figure: Pulse Measurement Tab Details. Only PG4 is enabled, and it is set for inverted polarity since a pulse modulator test set is being used.
Note 
If a noise floor signature is observed in the measurement, it is a possible indication that the pulse generator feeding the pulse modulator test set was not inverted.
The receiver channels remained coupled in this example, although only b2 is being used for these measurements.
Sweep 1 Settings in Pulse Configuration Dialog Box
Note 
The following receiver notations used in text and in the instrument user interface have identical meanings: A1 = a1, B1 = b1, A2 = a2, B2 = b2.
 
Sweep 2 Settings in the Pulse Configuration Dialog Box
Pulse Measurement Tab Details
Point-in-Pulse Example Measurement Results
The measurement results for the example described above are illustrated in Figure: Example Results of Two Point-in-Pulse Measurements.
There is a noticeable decrease in saturated power from this device as the duty cycle increases and the measurement pulse is repositioned (relative to the leading edge of the stimulus), as might be expected if there were significant thermal effects present. A receiver calibration was used for this measurement (performed with a simple thru line connection as discussed in other chapters in this guide).
Example Results of Two Point-in-Pulse Measurements
Only power and receiver calibrations are applied.