Sets the ‘eye amplitude’ display state to ON/OFF for the active trace of the indicated channel for NRZ signaling type. This corresponds to the Amplitude selection on the AMPLITUDE SETUP menu.
Returns the ‘eye amplitude’ display state for the active trace of the indicated channel for NRZ signaling type.
Sets the ‘eye duty cycle distortion’ display state to ON/OFF for the active trace of the indicated channel for NRZ signaling type. This corresponds to the Duty Cycle Distortion selection on the TIME SETUP menu.
Returns the ‘eye duty cycle distortion’ display state for the active trace of the indicated channel for NRZ signaling type.
Sets the ‘eye fall time’ display state to ON/OFF for the active trace of the indicated channel for NRZ signaling type. This corresponds to the Fall Time selection on the TIME SETUP menu.
Returns the ‘eye fall time’ display state for the active trace of the indicated channel for NRZ signaling type.
Sets the ‘eye height’ display state to ON/OFF for the active trace of the indicated channel for NRZ signaling type. This corresponds to the Height selection on the AMPLITUDE SETUP menu.
Returns the ‘eye height’ display state for the active trace of the indicated channel for NRZ signaling type.
Sets the X offset location of the Eye Statistics readout for the given trace on the indicated channel for NRZ signaling type.
Returns the X Offset location of the Eye Statistics readout for the given trace on the indicated channel for NRZ signaling type.
The location is the top left corner of the trace display area. A value in the range of 0-100 can be set which can suitably get coerced based on the display size of the trace.
Cmd Parameters
<NR3> The input parameter is a unitless number between 0-100.
Query Parameters
NA
Query Output
<NR3> The output parameter is a unitless number value between 0-100.
Sets the Y offset location of the Eye Statistics readout for the given trace on the indicated channel for NRZ signaling type.
Returns the Y Offset location of the Eye Statistics readout for the given trace on the indicated channel for NRZ signaling type.
The location is the top left corner of the trace display area. A value in the range of 0-100 can be set which can suitably get coerced based on the display size of the trace.
Cmd Parameters
<NR3> The input parameter is a unitless value between 0-100.
Query Parameters
NA
Query Output
<NR3> The output parameter is a unitless value between 0-100.
Sets the ‘eye level mean’ display state to ON/OFF for the active trace of the indicated channel for NRZ signaling type. This corresponds to the Level Mean selection on the AMPLITUDE SETUP menu.
Returns the ‘eye level mean’ display state for the active trace of the indicated channel for NRZ signaling type.
Sets the ‘eye opening factor’ display state to ON/OFF for the active trace of the indicated channel for NRZ signaling type. This corresponds to the Opening Factor selection on the AMPLITUDE SETUP menu.
Returns the ‘eye opening factor’ display state for the active trace of the indicated channel for NRZ signaling type.
Sets the ‘eye level 1’ display state for to ON/OFF for the active trace of the indicated channel for NRZ signaling type. This corresponds to the ‘1’ Level selection on the AMPLITUDE SETUP menu.
Returns the ‘eye level 1’ display state for the active trace of the indicated channel for NRZ signaling type.
Sets the ‘eye PP jitter’ display state to ON/OFF for the active trace of the indicated channel for NRZ signaling type. This corresponds to the Pk-Pk Jitter selection on the TIME SETUP menu.
Returns the ‘eye PP jitter’ display state for the active trace of the indicated channel for NRZ signaling type.
Sets the ‘eye RMS jitter’ display state to ON/OFF for the active trace of the indicated channel for NRZ signaling type. This corresponds to the RMS Jitter selection on the TIME SETUP menu.
Returns the ‘eye RMS jitter’ display state for the active trace of the indicated channel for NRZ signaling type.
Sets the ‘eye rise time’ display state to ON/OFF for the active trace of the indicated channel for NRZ signaling type. This corresponds to the Rise Time selection on the TIME SETUP menu.
Returns the ‘eye rise time’ display state for the active trace of the indicated channel for NRZ signaling type.
Sets the ‘eye signal to noise’ display state to ON/OFF for the active trace of the indicated channel for NRZ signaling type. This corresponds to the Signal to Noise selection on the AMPLITUDE SETUP menu.
Returns the ‘eye signal to noise’ display state for the active trace of the indicated channel for NRZ signaling type.
Sets the ‘eye width’ on/off display state for the active trace of the indicated channel for NRZ signaling type. This corresponds to the Width selection on the Time Setup menu.
Returns the ‘eye width’ display state for the active trace of the indicated channel for NRZ signaling type.
Sets the ‘eye level 0’ display on/off state for the active trace of the indicated channel for NRZ signaling type. This corresponds to the ‘0’ Level selection on the AMPLITUDE SETUP menu.
Returns the ‘eye level 0’ display state for the active trace of the indicated channel for NRZ signaling type.
Sets the ‘eye amplitude’ display on/off state for the active trace of the indicated channel for the indicated eye for PAM-4 signaling type. This corresponds to the Amplitude selection on the AMPLITUDE SETUP menu.
Returns the ‘eye amplitude’ display state for the active trace of the indicated channel for the indicated eye for PAM-4 signaling type.
Sets the display ON/OFF of the amplitude-related eye statistics for the active trace of the indicated channel for PAM-4 signaling type.
Returns the display state of the amplitude-related eye statistics for the active trace of the indicated channel for the indicated eye for PAM-4 signaling type.
Sets the ‘eye duty cycle distortion’ on/off display state for the active trace of the indicated channel for the indicated eye for PAM-4 signaling. This corresponds to the Duty Cycle Distortion selection on the TIME SETUP menu.
Outputs the ‘eye duty cycle distortion’ display state type for the active trace of the indicated channel for the indicated eye for PAM-4 signaling.
Sets the ‘eye fall time’ display state to ON/OFF for the active trace of the indicated channel for the indicated eye for PAM-4 signaling. This corresponds to the Fall Time selection on the TIME SETUP menu.
Returns the ‘eye fall time’ display state for the active trace of the indicated channel for the indicated eye for PAM-4 signaling.
Sets the ‘eye height’ display state to ON/OFF for the active trace of the indicated channel for the indicated eye for PAM-4 signaling. This corresponds to the Height selection on the AMPLITUDE SETUP menu.
Returns the ‘eye height’ display state for the active trace of the indicated channel for the indicated eye for PAM-4 signaling.
Sets the X offset location of the Eye Statistics readout for the given trace on the indicated channel for PAM-4 signaling. Returns the X Offset location of the Eye Statistics readout for the given trace on the indicated channel for PAM-4 signaling.
The location is the top left corner of the trace display area. A value in the range of 0-100 can be set which can suitably get coerced based on the display size of the trace.
Cmd Parameters
<NR3> The input parameter is a unitless number between 0-100.
Query Parameters
NA
Query Output
<NR3> The output parameter is a unitless number value between 0-100.
Sets the Y offset location of the Eye Statistics readout for the given trace on the indicated channel for PAM-4 signaling.
Returns the Y Offset location of the Eye Statistics readout for the given trace on the indicated channel for PAM-4 signaling.
The location is the top left corner of the trace display area. A value in the range of 0-100 can be set which can suitably get coerced based on the display size of the trace.
Cmd Parameters
<NR3> The input parameter is a unitless value between 0-100.
Query Parameters
NA
Query Output
<NR3> The output parameter is a unitless value between 0-100.
Sets the ‘eye level mean’ display state to ON/OFF for the active trace of the indicated channel for the indicated eye for PAM-4 signaling. This corresponds to the Level Mean selection on the AMPLITUDE SETUP menu.
Returns the ‘eye level mean’ display state for the active trace of the indicated channel for the indicated eye for PAM-4 signaling type.
Sets the ‘eye opening factor’ display state to ON/OFF for the active trace of the indicated channel for the indicated eye for PAM-4 signaling. This corresponds to the Opening Factor selection on the AMPLITUDE SETUP menu.
Returns the ‘eye opening factor’ display state for the active trace of the indicated channel for the indicated eye for PAM-4 signaling type.
Sets the high-level display state for to ON/OFF for the active trace of the indicated channel for the indicated eye for PAM-4 signaling. This corresponds to the ‘1’ Level selection on the AMPLITUDE SETUP menu.
Returns the low-level display state for the active trace of the indicated channel for the indicated eye for PAM-4 signaling.
Sets the ‘eye PP jitter’ display state to ON/OFF for the active trace of the indicated channel for the indicated eye for PAM-4 signaling. This corresponds to the Pk-Pk Jitter selection on the TIME SETUP menu.
Returns the ‘eye PP jitter’ display state for the active trace of the indicated channel for the indicated eye for PAM-4 signaling.
Sets the ‘eye RMS jitter’ display state to ON/OFF for the active trace of the indicated channel for the indicated eye for PAM-4 signaling. This corresponds to the RMS Jitter selection on the TIME SETUP menu.
Returns the ‘eye RMS jitter’ display state for the active trace of the indicated channel for the indicated eye for PAM-4 signaling.
Sets the ‘eye rise time’ display state to ON/OFF for the active trace of the indicated channel for the indicated eye for PAM-4 signaling. This corresponds to the Rise Time selection on the TIME SETUP menu.
Returns the ‘eye rise time’ display state for the active trace of the indicated channel for the indicated eye for PAM-4 signaling.
Sets thee ‘eye signal to noise’ display state to ON/OFF for the active trace of the indicated channel for the indicated eye for PAM-4 signaling. This corresponds to the Signal to Noise selection on the AMPLITUDE SETUP menu.
Returns the ‘eye signal to noise’ display state for the active trace of the indicated channel for the indicated eye for PAM-4 signaling.
Sets the ‘eye width’ display state to ON/OFF for the active trace of the indicated channel for the indicated eye for PAM-4 signaling. This corresponds to the Width selection on the TIME SETUP menu.
Returns the ‘eye width’ display state for the active trace of the indicated channel for PAM-4 signaling.
Sets the low-level display state to ON/OFF for the active trace of the indicated channel for the indicated eye for PAM-4 signaling. This corresponds to the ‘0’ Level selection on the AMPLITUDE SETUP menu.
Returns the low-level display state for the active trace of the indicated channel for the indicated eye for PAM-4 signaling.