VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :CALCulate{1-16}[:SELected]:PARameter{1-16}:MEMory Subsystem
 
:CALCulate{1-16}[:SELected]:PARameter{1-16}:MEMory Subsystem
The :CALCulate{1-16}[:SELected]:PARameter{1-16}:MEMory{1-20} subsystem allows users to configure multiple memory traces. Each channel may have up to 20 memory traces, which can be spread among the regular display traces. Each trace can display up to 20 memory traces.
Memory Subsystems
Related memory commands are described in the following subsystems:
:CALCulate{1-16}:PARameter{1-16}:MEMory and MEMory{1-20} Subsystem
:CALCulate{1-16}[:SELected]:MEMory{1-20} Subsystem
:CALCulate{1-16}[:SELected]:PARameter{1-16}:MEMory Subsystem
 
:CALCulate{1-16}[:SELected]:PARameter{1-16}:MEMory:ACTive?
Description
Query only.
Returns the active memory location for the specified trace and channel.
Cmd Parameters
NA
Query Parameters
NA
Query Output
<char> ML1 | ML2 | ML3 | ML4 | ML5 | ML6 | ML7 | ML8 | ML9 | ML10 | ML11 | ML12 | ML13 | ML14 | ML15 | ML16 | ML17 | ML18 | ML19 | ML20
Range
NA
Default Value
ML1 for Trace1, ML2 for Trace2, ... , ML16 for Trace16
Syntax Example
:CALC1:PAR1:MEM:ACT?
:CALCulate{1-16}[:SELected]:PARameter{1-16}:MEMory:MEMorize <char>
Description
Saves trace data for selected channel and trace to the specified trace memory location.
No query.
Cmd Parameters
<char> ML1 | ML2 | ML3 | ML4 | ML5 | ML6 | ML7 | ML8 | ML9 | ML10 | ML11 | ML12 | ML13 | ML14 | ML15 | ML16 | ML17 | ML18 | ML19 | ML20
Query Parameters
NA
Query Output
NA
Range
NA
Default Value
NA
Syntax Example
:CALC1:PAR1:MEM:MEM ML5