:SENSe{1-16}:CORRection:COLLect:FLEXible Subsystem
The :SENSe{1-16}:CORRection:COLLect:FLEXible command defines the S-Parameter set for the existing calibration type on the indicated channel.
Calibration Option Subsystems
Related calibration option configuration and control subsystems are:
:SENSe{1-16}:CORRection:COLLect:FLEXible:DEFine <char>{,<char2>,...,<charn>}
:SENSe{1-16}:CORRection:COLLect:FLEXible:DEFine?
Description
The use of S-Parameters with Port 3 and/or Port 4 requires a 4-port VNA instrument.
Defines the S-Parameter set for the existing calibration type on the indicated channel. At least one S-Parameter must be selected. Up to 16 S-Parameters may be selected in any sequence.
Returns the S-Parameter set defined for the existing calibration type on the indicated channel.
Cmd Parameters
<char>{,<char2>,...<char6>} S11 | S12 | S21 | S22 | S13 | S31 | S23 | S32 | S33 | S14 | S41 | S24 | S42 | S34 | S43 | S44
Query Parameters
<char>{,<char2>,...<char6>} S11 | S12 | S21 | S22 | S13 | S31 | S23 | S32 | S33 | S14 | S41 | S24 | S42 | S34 | S43 | S44
Range
NA
Default
S11
Syntax Example
:SENS1:CORR:COLL:FLEX:DEF S11, S12
:SENS1:CORR:COLL:FLEX:DEF?
:SENSe{1-16}:CORRection:COLLect:FLEXible[:STATe] <char>
:SENSe{1-16}:CORRection:COLLect:FLEXible[:STATe]?
Description
Sets on/off the flexible definition of a calibration on the indicated channel.
Returns the on/off status of the flexible definition of a calibration on the indicated channel.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
<char> 1 | 0
Range
NA
Default
0
Syntax Example
:SENS1:CORR:COLL:FLEX ON
:SENS1:CORR:COLL:FLEX?