VectorStar™ MS464xB Series Microwave Vector Network Analyzer User Interface Reference : IMDView™ : IMDView Setup and Configuration Menus
 
IMDView Setup and Configuration Menus
IMDVIEW SETUP Menu – Swept Frequency Mode
Previous
Application Menu Set for IMDView
Prerequisites
Option 44 installed on the VNA, and APPLICATION Menu Transmission/Reflection mode activated.
Navigation
MAIN | Application | APPLICATION | IMDView Setup | IMDVIEW SETUP
IMDView Setup Menu – Swept Frequency Mode
IMDView ON/OFF
Toggles the IMDView application on or off.
IMD Mode
Opens the IMD MODE SELECT menu. Screen shot to the left shows the menu when Swept Frequency mode is selected from CURRENT MODE menu (see Figure: CURRENT MODE Menu).
IMDView Configuration
Opens the IMDView CONFIGURATION dialog box.
Center Frequency Start
Sets the center frequency (fc) start.
Center Frequency Stop
Sets the center frequency (fc) stop.
Tone Delta
The spacing between the main tones. In other literature, it is sometimes called ‘offset’ or ‘tone spacing’.
Tone 1 Power
IMD measurement Tone 1 power level.
Tone 2 Power
IMD measurement Tone 2 power level.
IMDVIEW SETUP Menu – SpectrumView Mode
IMDView Setup Menu – SpectrumView Mode
IMDView ON/OFF
Toggles the IMDView Application on or off.
IMD Mode
Opens the IMD Mode Select menu. Screen shot to the left shows the menu when SpectrumView mode is selected from CURRENT MODE menu (see Figure: CURRENT MODE Menu).
IMDView Configuration
Opens the IMDVIEW CONFIGURATION dialog box.
Center Frequency
fc (Center frequency): In the IMD mode, it refers to the frequency half-way between the two main tones.
Tone Delta
The spacing between the main tones. In other literature, it is sometimes called ‘offset’ or ‘tone spacing’.
Tone 1 Power
IMD measurement Tone 1 power level.
Tone 2 Power
IMD measurement Tone 2 power level.
IMD MODE Menu (Sweep Modes)
Previous
IMDVIEW SETUP Menu – Swept Frequency Mode
Prerequisites
Option 44 installed on the VNA, and APPLICATION menu Transmission/Reflection mode activated
Navigation
MAIN | Application | APPLICATION | IMDView Setup | IMDVIEW SETUP | IMD Mode | IMD MODE SELECT
CURRENT MODE Menu
Swept Frequency
Swept frequency IMD mode causes the center frequency fc to sweep over the indicated range meaning the two tones from the sources are also sweeping. At each frequency ‘point’, the sources are stationary and the receiver will scan the tones and products as desired.
SpectrumView
In SpectrumView measurement mode the main tones are fixed and the receiver sweeps over some specified range to look at a maximum product order. There are two different SpectrumView sweep modes: Linear and Segmented.
IMDView CONFIGURATION Dialog Box
Previous
IMDVIEW SETUP Menu – SpectrumView Mode
Prerequisites
Option 32 and Option 44 installed on the VNA, and APPLICATION menu Transmission/Reflection mode activated.
Navigation
MAIN | Application | APPLICATION | IMDView Setup | IMDVIEW SETUP | IMDView Configuration | IMDVIEW CONFIGURATION
Figure: IMDView Configuration Dialog – Swept Frequency Sweep (Linear) illustrates the Swept Frequency IMD Mode dialog when Freq Sweep (Linear) is selected as the sweep type. Selecting the Response Configuration button opens Response Configuration dialog box shown in Figure: Response Configuration Dialog.
IMDView Configuration Dialog – Swept Frequency Sweep (Linear)
Response Configuration Dialog  
Variables for IMD measurement types shown above are described in Table: IMD Measurement Variables.
IMD Measurement Variables
Measurement
Valid Orders
Valid IM Side
Reference Plane Has Effect?
Reference Tone Has Effect?
Tone Power
All
All
Yes
No
IMD Relative to Carrier
All
(Except order 1 will be trivial or an asymmetry measurement)
All
No
Yes
Intercept Point
All except 1
All
Yes
Yes
Tone Gain
Only 1
Only lower frequency
No
No
Asymmetry
All
All except average
No
No
Figure: IMDView Configuration Dialog – SpectrumView Frequency Sweep (Linear) illustrates the Spectrum View IMD Mode dialog when Freq Sweep (Linear) is selected as the sweep type.
IMDView Configuration Dialog – SpectrumView Frequency Sweep (Linear)
Figure: IMDView Configuration Dialog – Dual Source with Ports 1 & 3 as Two Tones illustrates the settings for a 4-port VNA with Dual source (opt. 31) – port 1 and port 3 as two tones.
IMDView Configuration Dialog – Dual Source with Ports 1 & 3 as Two Tones
Figure: IMDView Configuration Dialog – Dual Source with Ports 2 & 4 as Two Tones illustrates the settings for a 4-port VNA with Dual source (opt. 31) – port 2 and port 4 as two tones.
IMDView Configuration Dialog – Dual Source with Ports 2 & 4 as Two Tones
Dialog Box Tone Selections vs VNA Configuration
Figure: IMDView Tone Selections vs. VNA Configuration – 2-port (1 of 3) lists available tone selections from the Tone 1 and Tone 2 dropdown menus in the IMDVIEW CONFIGURATION dialog, based on a 2-port VNA configuration.
IMDView Tone Selections vs. VNA Configuration – 2-port (1 of 3)
Standard Mode – Single Source VectorStar (no Option 31):
Int Src->Port 1
Int Src->Port 2
Ext Source 1
Ext Source 2
Ext Source 3
Ext Source 4
Standard Mode – Dual Source VectorStar (with Option 31):
Src 1->Port 1
Src 2->Port 1 (Opt 32 installed)
Src 2->Port 2
Ext Source 1
Ext Source 2
Ext Source 3
Ext Source 4
Broadband Mode – Single Source VectorStar (no Option 31) 54 GHz:
Int Src->Mod 1
Int Src->Mod 2
Ext Synth 1
Ext Synth 2
Ext Synth 3
Ext Synth 4
Broadband Mode – Dual Source VectorStar (with Option 31) 54 GHz:
Src 1->Mod 1
Src 2->Mod 1
Src 2->Mod 2
Ext Synth 1
Ext Synth 2
Ext Synth 3
Ext Synth 4
Broadband Mode – Single Source VectorStar (no Option 31) >54 GHz:
Int Src->Mod 1
Int Src->Mod 2
Ext Synth 1 -> Aux Mod
Ext Synth 2 -> Aux Mod
Ext Synth 3 -> Aux Mod
Ext Synth 4 -> Aux Mod
Broadband Mode – Dual Source VectorStar (with Option 31) >54 GHz:
Src 1->Aux Mod
Src 2->Mod 1 (Opt 32 installed)
Src 2->Mod 2
Ext Synth 1 -> Aux Mod
Ext Synth 2 -> Aux Mod
Ext Synth 3 -> Aux Mod
Ext Synth 4 -> Aux Mod
Figure: IMDView Tone Selections vs. VNA Configuration – 4-port (1 of 3) lists available tone selections from the Tone 1 and Tone 2 dropdown menus in the IMDVIEW CONFIGURATION dialog, based on a 4-port VNA configuration.
IMDView Tone Selections vs. VNA Configuration – 4-port (1 of 3)
Standard Mode – Single Source VectorStar (no Option 31):
Int Src->Port 1
Int Src->Port 2
Int Src->Port 3
Int Src->Port 4
Ext Source 1
Ext Source 2
Ext Source 3
Ext Source 4
Standard Mode – Dual Source VectorStar (with Option 31):
Src 1->Port 1
Src 1->Port 2
Src 2->Port 3
Src 2->Port 4
Ext Source 1
Ext Source 2
Ext Source 3
Ext Source 4
Broadband Mode – Single Source VectorStar (no Option 31) 54 GHz:
Int Src->Mod 1
Int Src->Mod 2
Int Src->Mod 3
Int Src->Mod 4
Ext Source 1
Ext Source 2
Ext Source 3
Ext Source 4
Broadband Mode – Dual Source VectorStar (with Option 31) 54 GHz:
Src 1->Mod 1
Src 1->Mod 2
Src 2->Mod 3
Src 2->Mod 4
Ext Source 1
Ext Source 2
Ext Source 3
Ext Source 4
Broadband Mode – Single Source VectorStar (no Option 31) >54 GHz:
Int Src->Mod 1
Int Src->Mod 2
Int Src->Mod 3
Int Src->Mod 4
ExtSrc1 -> AuxMod1
ExtSrc1 -> AuxMod2
ExtSrc2 -> AuxMod1
ExtSrc2 -> AuxMod2
ExtSrc3 -> AuxMod1
ExtSrc3 -> AuxMod2
ExtSrc4 -> AuxMod1
ExtSrc4 -> AuxMod2
Broadband Mode – Dual Source VectorStar (with Option 31) >54 GHz:
Src 1->Mod 1
Src 1->Mod 2
Src 2->Mod 1
Src 2->Mod 2
Src 3->Mod 1
Src 3->Mod 2
Src 4->Mod 1
Src 4->Mod 2
ExtSrc1 -> AuxMod1
ExtSrc1 -> AuxMod2
ExtSrc2 -> AuxMod1
ExtSrc2 -> AuxMod2
ExtSrc3 -> AuxMod1
ExtSrc3 -> AuxMod2
ExtSrc4 -> AuxMod1
ExtSrc4 -> AuxMod2