Setting Up Traditional Frequency Sweeps (Linear and Log)
A traditional frequency sweep is based on a start frequency, a stop frequency, and a number of points (or, alternatively, substitute center/span for start/stop). The number of points is not confined to certain preset values. The minimum number is two (otherwise use CW mode) and the maximum number is usually 25,000.
A mode allowing 100,000 points is also available, but operation is limited to a single channel.
Main POWER Menu—Port 1 and Port 2 Coupled—Attenuators Installed
Power Cal Menu
POWER CAL Menu
The objective of the power cal is to improve the accuracy of the power delivered to the DUT beyond that provided by the factory ALC calibration (0.1 dB vs. on the order of 1 dB). This is particularly useful if a preamplifier or other network is needed between the test port and the DUT. The exact loss/gain of that network over frequency can be corrected for with reasonable precision. A common setup for executing this calibration is shown in Figure: Port 1 Power Calibration Example Setup.
Port 1 Power Calibration Example Setup
The relationship between the power entry fields and the Target Power field of the power calibration is important. First a few definitions:
ALC Entry Field (Port ‘n’ Power on Power Menu):
Where the requested power is entered on the main Power menu for Port 1, Port 2, Src2 out Port 1, etc. (this will also apply to the >54 GHz power fields for the 3739-test-set-based mmWave modes and to the Aux Module power levels for mmWave IMD measurements (when present).
Effective Power Field (on Power Menu):
The read-only fields below the ALC entry fields. These normally reflect the settings of source attenuators (if present) but play an additional role when power calibrations are active.
Target Power Field (on Power Cal Menu):
The entry area on the Power Cal submenu where the objective of the power calibration is entered.
The normal procedure is to enter the desired power at the user reference plane (where the power sensor will be connected for the power calibration). If the gain/loss between the VNA port and the user reference plane is small, the ALC entry field can be set to the same value as the Target Power. If the gain/loss is significant, the ALC field should be offset by an estimate of that gain/loss (ALC entry field should be higher if it is a loss, and lower if it is a gain). This offset need not be extremely accurate; it serves to inform the system of where roughly to set the internal sources in order to achieve the desired user reference plane power. After the power calibration is completed, the Effective power field will match the Target power. If the ALC entry field is later changed (from the value at the time of the calibration), the Effective power field will also change to reflect an estimate of the new user reference plane power. This new value is only an estimate since the calibration was not performed at that power level and the accuracy will decrease as the change from the calibrated level increases.
Example
Target power is –20 dBm and there is about 8 dB of loss from the VNA port to the desired reference plane.
1. Set Target Power to –20 dBm and set the ALC entry field (for that port) to –12 dBm. Perform the power calibration.
2. After the calibration is successfully completed, the Effective power will read –20 dBm and the ALC entry field will remain at –12 dBm.
3. Change the ALC entry field to –15 dBm. The Effective power field will change to –23 dBm (the estimate of the new power at the reference plane).
4. Turn the power calibration off. If the ALC entry field is still at –15 dBm, the Effective power field will change to also be at –15 dBm. With the power calibration no longer applied, the system’s best estimate of the power being applied is that based on the internal power calibration.
Note that the source attenuator settings (if Option 61 or 62 is installed) will be reflected in the Effective Power value, even if no power calibration is applied.
Since the power cal performs the calibration at every point, this calibration can be time-consuming (particularly if a slower thermal power sensor is being used). One should exercise some restraint when selecting the number of power points if this time delay will be an issue. Details of power meter connection and setup are covered in the Operation Manual, but one should ensure that the power meter GPIB address matches that shown on the MS464xB Series VNA GPIB menu system and that the dedicated GPIB connector on the VNA is used. The dialog that appears when executing this calibration is shown in Figure: POWER CALIBRATION (PORT 1) Dialog Box.
POWER CALIBRATION (PORT 1) Dialog Box
Power Meter/Sensor Selection Options
Many possible power meters/sensors can be used with the MS464xB VNA and these choices proliferate with the broadband and millimeter wave versions (see Broadband/mmWave Measurements (Option 7, Option 8x) of this guide) so some guidance may be helpful:
• The ML24XX series of power meters and appropriate sensors can be used if the frequency ranges match (of the sensor and the range over which the calibration is being attempted) and these are usually the default. Control must be via GPIB and the address the VNA will look for is determined by the Ext. Power Meter address entry under System/Remote Interface/Power meters. The highest frequency sensor available in this family reaches 70 GHz and models are available that reach down to 70 kHz.
• Any GPIB meter can be used if it has a HP437B emulation mode. These meters will be expected also at the Ext. Power Meter address. Note that not all emulation modes are equal and the timing may not always behave correctly (particularly at very low power levels where settling time may be longer).
• Broadband (to 110 GHz) sensors are supported (also GPIB controlled, the R&S NRP2 or NRX meter with NRP-Z58 sensor as one example). This will be searched for under the Broadband sensor address.
• Certain waveguide-based sensors are specifically supported for millimeter wave applications.
• An HP437B-equivalent meter using a W8486A W-band sensor. This meter will be searched for under the W-band address.
• A D-band sensor (ELVA DPM-06 sensor with appropriate meter with GPIB) can be used above 110 GHz (for ME7838D/G systems mainly) and this is searched for under the D-band address. Normally the W8486A sensor will be used up to 125 GHz if it is available and the D-band sensor above that. The D-band sensor will be used above 110 GHz if the W8486A sensor is not available. Particularly for ME7838G systems, make sure an appropriate waveguide-to-0.6mm adapter is being used in that case (A WR6G adapter is available that covers the full 110-170 GHz range. The WR5G adapter can be used as well over most of this range as its cutoff frequency is 115 GHz).
• A G-band sensor (ELVA DPM-05 sensor with appropriate meter with GPIB) can be used for ME7838G systems mainly. This sensor is normally used only above 170 GHz but will be used above 140 GHz if a D-band sensor is not available and the calibration requires that frequency range. A 35WR5G adapter is normally used when using this sensor with ME7838G systems.
• Keysight power meters are also supported. The Keysight N191XX EPM series of meters in their native mode are supported under two different address classes: Ext. Power Meter (in lieu of the ML243XX power meter) and the W-band Power Meter (using the W8486A sensor). Since this meter supports sensors that cannot return their frequency limits, it is incumbent upon the user to only connect this meter when the sensor range is valid for the calibration of interest. The VNA will assume the sensor can cover 70 kHz–70 GHz if the meter is connected on the Ext. Power Meter address, and will assume it can cover the extended W-band if connected under the W-band Power Meter address.
• USB sensors from Anritsu are also supported where the sensor’s specific frequency range relative to the intended range for calibration is relevant. Multiple USB sensors can be selected and the user will be asked to choose at calibration time.
• Note: Usage of the MA24500A series sensor requires a dual USB Type A male to single USB Type A female cable to supply needed current draw.
• The MA25400A sensors are different from the sensors discussed above in that they use a heterodyne detection scheme rather than a broadband approach. The way this sensor is used in Vector Star calibrations, the integration bandwidth is 7 MHz, and this has some implications. First, the minimum detectable power is in the range of -35 dBm (it can be lower when the sensor is used with a separate PC when the integration bandwidth can be smaller). Second, the accuracy is slightly degraded at very low frequencies (principally below 500 kHz) when the front end corrections start deviating due to the bandwidth.
When not in broadband mode, the user will have the option to select a USB sensor if one is detected. If not, the system will first scan for the Ext. Power Meter address and will use that meter if found and the reported frequency ranges match (note that 437B emulating meters do not report a frequency range so the system will use them regardless). If such a meter is not found, it will next search for the Broadband meter. If that is also not found, the calibration will terminate with an error.
In broadband modes, there are many more choices and the system will switch between sensors/meters to cover the desired range. This is perhaps illustrated best with a series of examples and those are in the chart shown in Table: Power Sensor Selection Guide (1 of 2). The system will start from the left of the chart and use the first option (or combination) that is available. Many of the sweep ranges in the chart are only possible with the ME7838x broadband systems (see Broadband/mmWave Measurements (Option 7, Option 8x) of this guide for more information).
1. Since no USB sensor is available, it skips the first device column.
2. A broadband and a W-band sensor are available so it will use the 2nd device column
The general principles are, in this order:
• If a USB sensor is connected and selected, use it as far as it can be used.
• Use the connected sensor that can cover as much of the frequency range as possible.
• Use the connected sensor with the lowest uncertainty.
One can always override this decision tree by only leaving connected the meters/sensors that one wants to use.
Power Sensor Selection Guide (1 of 2)
When Broadband is Selected
Power Cal Frequency Range
If USB Sensor Present and Selected
Next Choice
Next Choice After That
Next Choice After That
70 GHz CW
USB
Broadband
Anritsu GPIB or 437B emulation
W-Band
110 GHz CW
USB
Broadband
W-Band
D-Band
125 GHz CW
USB
W-Band
D-Band
10 MHz to 70 GHz
USB
Broadband
Anritsu GPIB or 437B emulation
10 MHz to 110 GHz
USB
Broadband
Anritsu GPIB or 437B emulation & W-Band
10 MHz to 125 GHz
USB
Broadband & W-Band
Anritsu GPIB or 437B emulation & W-Band
Broadband & D-Band
10 MHz to 150 GHz
USB
Broadband & W-Band & D-Band
Anritsu GPIB or 437B emulation & W-Band & D-Band
Broadband & D-Band
70 GHz to 110 GHz
USB
Broadband
W-Band
70 GHz to 125 GHz
USB
W-Band
Broadband & D-Band
70 GHz to 150 GHz
USB
W-Band & D-Band
Broadband & D-Band
110 GHz to 125 GHz
USB
W-Band
D-Band
110 GHz to 150 GHz
USB
W-Band & D-Band
D-Band
125 GHz to 150 GHz
USB
D-Band
Whatever the basic power meter accuracy is will be transferred to the VNA within about 0.1 dB (convergence criteria is normally 0.07 dB, this limit loosens for frequencies above 70 GHz) for short periods of time. That correlation can drift as temperatures change, connection repeatability, etc. The basic accuracy of the power meter is outside the scope of this guide (consult the documentation for the meter/sensor in question) but includes terms such as linearity, mismatch, basic power factor correction, offsets, and other items.
Embedding/De-embedding
An additional function possible with power calibrations is embedding or de-embedding where an existing power calibration is modified by the S-parameters in a .s2p file to reflect a network that was added after the power calibration (embedding) or removed after calibration (de-embedding). These functions are placed on a submenu labeled NW extraction (short for network extraction, this refers to the calibration options submenu where the .s2p file is often generated). They are accessed by selecting Power | Power Cal | NW Extraction. Figure: NW EXTRACTION Menu shows the NW Extraction menu, which is enabled only if a power cal exists.
NW EXTRACTION Menu
This menu enables one to load files separately for embedding and de-embedding. Note that the files are separate for port 1 and port 2 power calibrations. Also note that only the S21 parameter is used in the adjustment of the power calibration (and, of course, only the magnitude). Interpolation and extrapolation are used if the frequency lists of the .s2p file and that of the power calibration do not match. A warning dialog will appear if extrapolation will be used.
If the network loaded is lossy and is being embedded (e.g., the power calibration is performed coaxially and then the reference plane is connected to a wafer probe and the desired reference plane is at the probe tip), the power delivered by the source will increase by the amount of the loss. Similarly, if such a lossy network is de-embedded (e.g., an adapter was used to facilitate the calibration but is then removed for the measurement), the power delivered by the source will decrease by the amount of the loss.
All of the above discussion applies to both linear and log versions of swept frequency. In log sweep, the frequency list is calculated using a constant ratio step as opposed to a constant step size in linear frequency sweep. Also, the graticule for rectilinear plots while in log sweep will change to reflect that modified frequency list. An example is shown in Figure: Example Graticule for the Log Sweep Type. Integral numbers of decades of lines will always be used.