VectorStar™ MS464xB Series Microwave Vector Network Analyzer Programming Manual : SCPI Commands : :CALCulate{1-16}[:SELected]:SMOothing Subsystem
 
:CALCulate{1-16}[:SELected]:SMOothing Subsystem
The :CALCulate{1-16}[:SELected]:SMOothing subsystem commands are used to configure and control trace smoothing functions.
Trace Subsystems
Related trace subsystems are:
• :CALCulate{1-16}:PARameter and :PARameter{1-16} Subsystem
• :CALCulate{1-16}:PARameter{1-16}:FSIMulator Subsystem
• :CALCulate{1-16}:PARameter{1-16}:SELect Subsystem
• :CALCulate{1-16}:PARameter{1-16}:COUPling Subsystem
• :CALCulate{1-16}:POLar Subsystem
• :CALCulate{1-16}:PROCessing:ORDer Subsystem
• :CALCulate{1-16}[:SELected]:CONVersion Subsystem
• :CALCulate{1-16}[:SELected]:DATA Subsystem
• :CALCulate{1-16}[:SELected]:FORmat Subsystem
• :CALCulate{1-16}[:SELected]:MATH Subsystem
• :CALCulate{1-16}[:SELected]:MDATA Subsystem
• :CALCulate{1-16}[:SELected]:SMITh Subsystem
• :CALCulate{1-16}[:SELected]:SMOothing Subsystem
• :CALCulate{1-16}[:SELected]:TRACE{1-16}:MARKer{1-12} Subsystem
• :CALCulate{1-16}[:SELected]:TDATA Subsystem
• :DISPlay Subsystem
:CALCulate{1-16}[:SELected]:SMOothing:APERture <NRf>
:CALCulate{1-16}[:SELected]:SMOothing:APERture?
Description
Sets the smoothing aperture for the indicated channel and active trace.
Returns the smoothing aperture for the indicated channel and active trace.
Cmd Parameters
<NRf> The input parameter is in Percent.
Query Parameters
<NR3> The output parameter is in Percent.
Range
0 to 100
Default Value
0.000000E+000
Syntax Example
:CALC1:SMO:APER 2
:CALC1:SMO:APER?
:CALCulate{1-16}[:SELected]:SMOothing[:STATe] <char>
:CALCulate{1-16}[:SELected]:SMOothing[:STATe]?
Description
Sets the smoothing on/off state for the indicated channel and active trace.
Returns the smoothing on/off status for the indicated channel and active trace.
Cmd Parameters
<char> 1 | 0 | ON | OFF
Query Parameters
<char> 1 | 0
Range
NA
Default Value
0
Syntax Example
:CALC1:SMO ON
:CALC1:SMO?