Cable and Antenna Analyzer Measurement Guide : Cable and Antenna Analyzer : Cable and Antenna Measurement Setup : Frequency
 
Frequency
For VSWR, Return Loss, Cable Loss, 2-Port Gain, Smith Chart, 1-Port Phase, and 2-Port Phase measurements.
Setting up the Measurement Frequency using Start and Stop Frequencies
1. Press the Freq/Dist main menu key.
2. Press the Start Freq submenu key and use the keypad to enter the start frequency.
When entering a frequency using the keypad, the key labels change to GHz, MHz, kHz, and Hz. Press the appropriate unit key to complete the entry.
3. Press Stop Freq and use the keypad to enter the stop frequency. Press the appropriate unit key to complete the entry.
Setting up the Measurement Frequency by Selecting a Signal Standard
1. Press the Freq/Dist main menu key.
2. Press the Signal Standard submenu key.
3. Select uplink, downlink, or uplink plus downlink.
4. Press the Select Standard key.
5. Use the rotary knob or the Up/Down arrow keys and scroll to the appropriate signal standard and press Enter to select.
Note 
The Signal Standard menu can be customized. If a particular standard is missing, Line Sweep Tools (LST) can be used to edit the signal standard list. Please refer to the LST Help menu for details.
Frequency/Distance
(Distance-To-Fault Return Loss, Distance-To-Fault VSWR)
1. Press the Freq/Dist main menu key.
2. Press the Units key to toggle between using meter (m) or feet (ft).
3. Press the Start Dist submenu key and use the arrow keys, rotary knob, or keypad to enter the start distance.When entering a distance using the keypad, press the unit key or Enter to complete the entry.
Note 
The Start Dist cannot be set beyond the Stop Dist.
4. Press Stop Dist and enter the stop distance.
5. To set the frequency, press DTF Aid. For more details about DTF Aid, refer to DTF Measurement.
Refer to Freq/Dist Menu for additional information.