PIM Master™ Passive Intermodulation Analyzer with Site Master™ Cable & Antenna Analyzer (331) and 2-Port Options (703) User Guide : Instrument Care
 
Instrument Care
Introduction
This appendix provides instructions for preventive maintenance, battery care, and firmware updates.
Appendix Overview
This appendix contains the following sections:
Preventive Maintenance
Battery Care
Battery Replacement
PIM Master Firmware Update
Adding Options with a License File
Preventive Maintenance
PIM Master preventive maintenance consists of cleaning the unit and inspecting and cleaning the RF connectors on the instrument and all RF cables. Clean the PIM Master with a soft, lint‑free cloth slightly dampened with water, or water and a mild cleaning solution.
Clean the RF connectors and center pins with a lint‑free wipe or cotton swab dampened with isopropyl alcohol. Use a non‑metallic object (such as the wooden end of a cotton swab) to push the lint‑free wipe into the connector RF interface to remove metal flakes. Visually inspect the connectors. The pins of the connectors should be unbroken and uniform in appearance. If you are unsure whether the connectors are undamaged, gauge the connectors to confirm that the dimensions are correct.
Visually inspect the test port cables. To obtain accurate readings within instrument specifications, test port cables must be: free of metal flakes, uniform in appearance, and not stretched, kinked, dented, or broken.