Vision™ MX280001A Software Help : Trace Monitor/Trace Viewer : Trace Matrix
 
Trace Matrix
The user can observe interference signal patterns that are occurring based on time and day. In Figure: Trace Matrix with One Trace Per Block, each block is 5 seconds and a row a tenth of an hour. Depending on the length of time measurements are taken, the amount of traces can fill the matrix display. If that occurs, use the vertical scroll bar of the display to view the other blocks of measurements.
Color Bars
The Color Bars button allows you to change the color scheme of the data blocks in the matrix from green, red, and yellow to blue, yellow, and green.
 
Trace Matrix with One Trace Per Block
For example, data collection is set to 10 seconds and the time frame in Vision Monitor is set to 6 hours making a data block 20 seconds long. This results in acquiring two traces for each data block and the color coding shown in Figure: Trace Matrix with Multiple Traces per Block.
Click on a colored square to view the associated trace in the Trace Panel or Spectrogram. Depending on how you set up data collection, multiple traces can be collected during one data block. The color definitions are:
Green: All traces pass in that defined block.
Red: All traces fail in that defined block.
Yellow: Some traces pass and some fail in the defined block.
Blank: Occurs during tracking events when Trace Acquire is not collecting traces and is determining the tracking candidate status.
Trace Matrix with Multiple Traces per Block