Vision™ MX280001A Software Help : Trace Monitor/Trace Viewer : Trace Mask : Trace Mode
 
Trace Mode
The Trace Mode box shows a list as shown in Figure: Trace Mode Menu.
Trace Mode Menu
Select the Trace Mode Setting:
Normal
Displays data for the current trace sweep. Can be applied to several channel monitors in the database.
Rolling Min Hold
Minimum value at each frequency point over the last trace sweeps. Can be applied to several channel monitors in the database.
Rolling Max Hold
Maximum value at each frequency point over the last trace sweeps. Can be applied to several channel monitors in the database.
Average
Average value at each frequency point over the last trace sweeps. Can be applied to several channels in the database. When set, the spectrum monitor sweeps and averages the set number of sweeps before it returns a trace.
Min Hold
Minimum value at each frequency point over the last trace sweep. Applied to only one channel monitor in the database. This returns what is in the Min Hold buffer on the spectrum analyzer. The time allocated to sweep takes the same time as a single trace sweep.
Max Hold
Maximum value at each frequency point over the last trace sweep. Applied to only one channel monitor in the database. This returns what is in the Max Hold buffer on the spectrum analyzer. The time allocated to sweep takes the same time as a single trace sweep.