Sets the maximum power level when level limit state is enabled. Range is -130 dBm to 33 dBm with Option 2. Range is –20 to 33 dBm without Option 2.
RF Level Limit State
When ON, the level status area displays “Limit” or “Lim”.
Retrace RF
Controls the RF output On/Off during retraces. Toggling retrace RF Output protection On turns RF Off during retrace.
Delta-F RF
Toggle that controls RF On or Off during frequency switching in CW, Step sweep, and list sweep modes. Toggle Retrace RF Output protection On turns Delta-F RF Off.
Units
Level units, select dBm, dBµV, mV.
Setting RF LEVEL LIMIT
RF LEVEL LIMIT sets the maximum power level when RF Level Limit State is enabled. The RF limit shown in Figure: RF Level Limit State shows the RF Limit Level State toggled on and the maximum achievable RF power level limit clamped to +5 dBm. In this example, the power level settings may be entered at a higher power level value, but the power output will remain at +5dBm until the RF Level Limit State is toggled off.
Press: LEVEL | SETUP | RF LEVEL LIMIT
RF Level Limit State
Setting Retrace RF
Switch Retrace RF to ON sets the RF output off during sweep retrace.
Switch Retrace RF to OFF sets the RF output on.
Setting Delta-F RF
Switch Delta-F RF to select RF on or off during frequency switching in CW, step sweep, and list sweep modes.
Units
Press Units to show the power level units as shown in Figure: LEVEL SETUP Units Menu. When SWEEP TYPE Log as described in Setting SWEEP TYPE is selected, the units are displayed in dBm or dBμV. When SWEEP TYPE Linear is selected, the units are displayed in mV. The units are displayed with the indicated power level.