SETTINGS RESET FACTORY RESET Restores the instrument to the factory default settings for all measurement modes and system settings, including language, display and audio settings. Ethernet and, time zone settings are not affected. The instrument will automatically restart. If the unit is connected to Wi-Fi, the previous connections will be preset. FILE RESET DELETE USER FILES Deletes all user files from the instrument’s internal memory, including measurement, setup, and screenshot files. System files are not affected. DELETE SYSTEM FILES Deletes all user system files from the instrument’s internal memory, including keyboard EZ names, cable and antennas, and log files. Other user files are not affected. RESET ALL RESET ALL Performs a Factory Reset as described above and deletes last saved settings, log files and user files from the instrument’s internal memory. The RESET ALL function overwrites all information stored in the non-volatile memory with default values. This includes the table of 2000 frequency sets used for the sweep mode and the five power-offset tables used for the user level flatness correction function. The sweep is defined by a start index and stop index. The instrument will automatically restart. |
Note | Resetting the instrument clears the current setup parameters. If these parameter values are needed for future testing, save them as a stored setup before resetting the signal generator. |
Model Number | Frequency Parameters (GHz) | |||||||||
F0 | F1 | F2 | F3 | F4 | F5 | F6 | F7 | F8 | F9 | |
MG36221A | 3.5 | 2.2 | 20.0 | 2.2 | 5.0 | 8.0 | 11.0 | 14.0 | 17.0 | 20.0 |
MG36241A | 3.5 | 2.2 | 43.5 | 2.2 | 5.0 | 8.0 | 11.0 | 14.0 | 17.0 | 20.0 |
MG36271A | 3.5 | 2.2 | 70.0 | 2.2 | 5.0 | 8.0 | 11.0 | 14.0 | 17.0 | 20.0 |
Model Number | Parameters (dBm) | |||||||||
L0 | L1 | L2 | L3 | L4 | L5 | L6 | L7 | L8 | L9 | |
MG362x1A | +1.0 | 0.0 | –1.0 | –2.0 | –3.0 | –4.0 | –5.0 | –6.0 | –7.0 | –8.0 |
Model Number | Sweep Time | Step Sweep | Level Sweep | Level Offset | ||
Dwell Time | Number of Points | Dwell Time | Number of Steps | |||
MG362x1A | 50 ms | 1 ms | 1000 | 1 ms | 1000 | 0.0 dB |