Most VNA calibrations are performed referenced to 50 ohms as this is usually set by the calibration kit. While some calibration kits exist for other impedances (75 ohm N and F connectors for example), they are not common and a custom impedance may be of interest. The impedance transformation function allows performance of a calibration in one impedance and then transform the result to appear as if it had been calibrated in a different impedance. As a crude example, if a 75 ohm N calibration kit is not available, but a 50 ohm kit is, neglecting issues with adapters, the 50 ohm calibration could be performed and the utility used to reference the results to 75 ohms. The
IMPED. TRANS. (
IMPEDANCE TRANSFORMATION) menu for this function is shown below (
Figure: IMPED. TRANSF. (IMPEDANCE TRANSFORMATION) Menu).
A calibration must be active for impedance transformation to take effect and only those ports currently calibrated will be affected.