VectorStar™ MS4640A SeriesMicrowave Vector Network Analyzer : SCPI Commands : :SENSe{1-16}:CORRection:COEFficient Subsystem - Simulation
 
:SENSe{1-16}:CORRection:COEFficient Subsystem - Simulation
The :SENSe{1-16}:CORRection:COEFficient subsystem commands are used to simulate a instrument calibration on 2-Port VNA instruments so that a set of user-defined calibration parameters can be input using the :SENSe{1-16}:CORRection:COEFficient <char1>,<block> command. Note that this subsystem does not perform an actual calibration.
Calibration Simulation Subsystems
These subsystems are used to create a calibrated state in the instrument which is followed by adding the required error correction coefficients for the required calibration type. If this approach is used, each error correction coefficient is entered by separate commands. Simulated calibration subsystems are:
:CALCulate{1-16}:FSIMulator:NETWork Subsystem - Simulation
:CALCulate{1-16}:FSIMulator:NETWork {1-50} Subsystem - Simulation
:CALCulate{1-16}:FSIMulator:NETWork {1-50} Subsystem - Simulation
:SENSe{1-16}:CORRection:COEFficient:PORT Subsystem - Simulation
:SENSe{1-16}:CORRection:COEFficient Subsystem - Simulation
Calibration Type Abbreviations
The calibration abbreviations and their calibration types are:
:1P2PF refers to a one-path two-port calibration forward direction
:1P2PR refers to a one path two port calibration reverse direction
:FULL1 refers to a full one port calibration
:FULL2 refers to a full two port calibration
:FULL3 refers to a full three port calibration
:FULLB refers to a full one port reflection calibration on both ports
:FULL4 refers to a full four port calibration
:RESP1 refers to a one port response calibration
:RESPB refers to a one port response calibration both ports
:TFRB refers to a transmission frequency response calibration both directions
:TFRF refers to a transmission frequency response calibration forward direction
:TFRR refers to a transmission frequency response calibration reverse direction
Each calibration simulation type command is described in greater detail in the individual command descriptions below.
These commands set the Calibration Type. To query which Calibration Type has been sect, use the command:
:SENSe{1-16}:CORRection:COLLect:TYPe?
 
:SENSe{1-16}:CORRection:COEFficient:1P2PF
Description
The command simulates a one-path two-port calibration forward direction on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-Port VNA or a 4-Port VNA instrument. No query. To query the state of this command use:
:SENSe{1-16}:CORRection:COLLect:TYPe?
To set this calibration mode and then perform an actual calibration, use:
:SENSe{1-16}:CORRection:COLLect:1P2PF
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COEF:1P2PF
 
:SENSe{1-16}:CORRection:COEFficient:1P2PR
Description
The command simulates a one-path Two-Port Calibration Reverse Direction on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-Port or 4-Port VNA instrument. No query. To query the state of this command, use:
:SENSe{1-16}:CORRection:COLLect:TYPe?
To set this calibration mode and then perform an actual calibration, use:
:SENSe{1-16}:CORRection:COLLect:1P2PR
No query.
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COEF:1P2PR
 
:SENSe{1-16}:CORRection:COEFficient:FULL1
Description
The command simulates a full One-Port Reflection Calibration on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-Port VNA or a 4-Port VNA instrument. No query. Before sending this command, the simulation port must be specified using:
:SENSe{1-16}:CORRection:COLLect:PORT
To query the state of this command, use:
:SENSe{1-16}:CORRection:COLLect:TYPe?
To set this calibration mode and then perform an actual calibration, use:
:SENSe{1-16}:CORRection:COLLect:FULL1
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COEF:FULL1
 
:SENSe{1-16}:CORRection:COEFficient:FULL2
Description
The command simulates a full Two-Port Calibration on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-Port or 4-Port VNA instrument. No query. To query the state of this command, use:
:SENSe{1-16}:CORRection:COLLect:TYPe?
To set this calibration mode and then perform an actual calibration, use:
:SENSe{1-16}:CORRection:COLLect:FULL2
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COEF:FULL2
 
:SENSe{1-16}:CORRection:COEFficient:FULLB
Description
The command simulates a full One-Port Reflection Calibration on both ports on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-Port or 4-Port VNA instrument. No query. To query the state of this command, use:
:SENSe{1-16}:CORRection:COLLect:TYPe?
To set this calibration mode and then perform an actual calibration, use:
:SENSe{1-16}:CORRection:COLLect:FULLB
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COEF:FULLB
 
:SENSe{1-16}:CORRection:COEFficient:RESP1
Description
The command simulates a One-Port Response Calibration on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-Port or 4-Port VNA instrument. No query.
Before sending this command, the simulation port must be specified using:
:SENSe{1-16}:CORRection:COLLect:PORT
To query the state of this command, use:
:SENSe{1-16}:CORRection:COLLect:TYPe?
To set this calibration mode and then perform an actual calibration, use:
:SENSe{1-16}:CORRection:COLLect:RESP1
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COEF:RESP1
 
:SENSe{1-16}:CORRection:COEFficient:RESPB
Description
The command simulates a One-Port Response Calibration on both ports on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-Port or 4-Port VNA instrument. No query. To query the state of this command, use:
:SENSe{1-16}:CORRection:COLLect:TYPe?
To set this calibration mode and then perform an actual calibration, use:
:SENSe{1-16}:CORRection:COLLect:RESPB
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COEF:RESPB
 
:SENSe{1-16}:CORRection:COEFficient:TFRB
Description
The command simulates a Transmission Frequency Response Calibration in both directions on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-Port or 4-Port VNA instrument. No query. To query the state of this command, use:
:SENSe{1-16}:CORRection:COLLect:TYPe?
To set this calibration mode and then perform an actual calibration, use:
:SENSe{1-16}:CORRection:COLLect:TFRB
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COEF:TFRB
 
:SENSe{1-16}:CORRection:COEFficient:TFRF
Description
The command simulates a Transmission Frequency Response Calibration in the forward direction on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-Port or 4-Port VNA instrument. No query. To query the state of this command, use:
:SENSe{1-16}:CORRection:COLLect:TYPe?
To set this calibration mode and then perform an actual calibration, use:
:SENSe{1-16}:CORRection:COLLect:TFRF
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COEF:TFRF
 
:SENSe{1-16}:CORRection:COEFficient:TFRR
Description
The command simulates a Transmission Frequency Response Calibration in the reverse direction on the indicated channel. The port pair is limited to Ports 1-2 but can be used on either a 2-Port or 4-Port VNA instrument. No query. To query the state of this command, use:
:SENSe{1-16}:CORRection:COLLect:TYPe?
To set this calibration mode and then perform an actual calibration, use:
:SENSe{1-16}:CORRection:COLLect:TFRR
Cmd Parameters
NA
Range
NA
Default Value
NA
Syntax Example
:SENS1:CORR:COEF:TFRR