VectorStar™ MS4640A SeriesMicrowave Vector Network Analyzer : SCPI Commands : :SENSe{1-16}:CORRection:COLLect:LRL:CALB Subsystem
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB Subsystem
The :SENSe{1-16}:CORRection:COLLect:LRL:CALB subsystem commands provide control of Line-Reflect-Line second calibration (or CALB) configuration parameters, execution, and output reporting. The :LRL:CALB commands require a 4-Port VNA instrument.
LRL Calibration Subsystems
The LRL-related calibration commands are organized into five (5) subsystems in the following sections:
:SENSe{1-16}:CORRection:COLLect:LRL:CALB Subsystem (this subsystem)
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-4} Subsystem
:SENSe{1-16}:CORRection:COLLect:LRL:PORT Subsystem - 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:LRL:SINGleton Subsystem - 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa] Subsystem
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND:COUNt <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND:COUNt?
Description
This command requires a 4-Port VNA instrument. The command sets the number of bands to use in the LRL CALB calibration on the given channel. The query outputs the number of bands to use in the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is an integer.
Query Parameters
<NR1> The output parameter is an integer.
Range
1 or 2
Default
1
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:BAND:COUN 2
:SENS1:CORR:COLL:LRL:CALB:BAND:COUN?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND1:REFLection:TYPe <char1>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND1:REFLection:TYPe?
Description
This command requires a 4-Port VNA instrument. The command sets the Band1 reflection type of the LRL CALB calibration on the given channel where the type can be set as:
OPENlike = The device type is more like an Open.
SHORTlike = The device type is more like a Short.
BOTH = The device can be considered to be both an Open and a Short.
The query outputs the Band1 reflection type of the LRL CALB calibration on the given channel.
Cmd Parameters
<char1> OPENlike | SHORTlike | BOTH
Query Parameters
<char1> OPEN | SHORT | BOTH
Range
NA
Default
OPEN
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:BAND1:REFL:TYP OPENlike
:SENS1:CORR:COLL:LRL:CALB:BAND1:REFL:TYP?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND2:REFLection:TYPe <char1>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND2:REFLection:TYPe?
Description
This command requires a 4-Port VNA instrument. The command sets the Band2 reflection type of the LRL CALB calibration on the given channel where the type can be set as:
OPENlike = The device type is more like an Open.
SHORTlike = The device type is more like a Short.
BOTH = The device can be considered to be both an Open and a Short.
The query outputs the band2 reflection type of the LRL CALB calibration on the given channel.
Cmd Parameters
<char1> OPENlike | SHORTlike | BOTH
Query Parameters
<char1> OPEN | SHORT | BOTH
Range
NA
Default
OPEN
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:BAND2:REFL:TYP SHORTlike
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:BAND2:REFLection:TYPe?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:LINE
Description
This command requires a 4-Port VNA instrument. The command initiates collection of the indicated Device Line Standard data for the CALB calibration on the indicated channel. No query.
Cmd Parameters
NA
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:LINE
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:LINE:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:LINE:FREQuency?
Description
This command requires a 4-Port VNA instrument. The command sets the reference frequency for loss of the given device of the LRL CALB calibration on the given channel. The query outputs the reference frequency for loss of the given device of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:LINE:FREQ 1.0E7
:SENS1:CORR:COLL:LRL:CALB:DEV1:LINE:FREQ?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:LINE:LENGth <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:LINE:LENGth?
Description
This command requires a 4-Port VNA instrument. The command sets the line length of the given device of the LRL CALB calibration on the given channel. The query outputs the line length of the given device of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:LINE:LENG 2.5E-2
:SENS1:CORR:COLL:LRL:CALB:DEV1:LINE:LENG?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:LINE:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:LINE:LOSS?
Description
This command requires a 4-Port VNA instrument. The command sets the line loss of the given device of the LRL CALB calibration on the given channel. The query outputs the line loss of the given device of the LRL CALB calibration on the given channel.
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:LINE:LOSS?
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
<NR3> The output parameter is in dB/mm.
Range
MPND/1000 (MPND divided by 1000)
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:LINE:LOSS 3.0E0
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:LINE:LOSS?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:C0 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:C0?
Description
This command requires a 4-Port VNA instrument. The command sets the capacitance of the match device on the given port of the LRL CALB calibration on the given channel. The query outputs the capacitance of the match device on the given port of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Farads.
Query Parameters
<NR3> The output parameter is in Farads.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:C0 3.01E-12
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:C0?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:C1 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:C1?
Description
This command requires a 4-Port VNA instrument. Sets the C1 coefficient of the match device on the given port of the LRL CALB calibration on the given channel. Outputs the C1 coefficient of the match device on the given port of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz.
Query Parameters
<NR3> The output parameter is in Farads/Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:C1 2.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:C1?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:C2 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:C2?
Description
This command requires a 4-Port VNA instrument. Sets the C2 coefficient of the match device on the given port of the LRL CALB calibration on the given channel. Outputs the C2 coefficient of the match device on the given port of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^2.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^2.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:C2 2.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:C2?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:C3 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:C3?
Description
This command requires a 4-Port VNA instrument. Sets the C3 coefficient of the match device on the given port of the LRL CALB calibration on the given channel. Outputs the C3 coefficient of the match device on the given port of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^3.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^3.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:C3 2.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:C3?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:L0 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:L0?
Description
This command requires a 4-Port VNA instrument. The command sets the inductance of the match device on the given port of the LRL CALB calibration on the given channel. Use of Port 3 or Port 4 requires a four-port instrument. The query outputs inductance of the match device on the given port of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Henrys.
Query Parameters
<NR3> The output parameter is in Henrys.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:L0?
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:L0 2.0E-6
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:L1 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:L1?
Description
This command requires a 4-Port VNA instrument. Sets the L1 coefficient of the match device on the given port of the LRL CALB calibration on the given channel. Outputs the L1 coefficient of the match device on the given port of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz.
Range
MPND
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:L1 1.4
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:L1?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:L2 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:L2?
Description
This command requires a 4-Port VNA instrument. Sets the L2 coefficient of the match device on the given port of the LRL CALB calibration on the given channel. Outputs the L2 coefficient of the match device on the given port of the LRL CALB calibration on the given channel
Cmd Parameters
<NRf> in units of Meters/Hertz^2 (Meters per Hertz squared)
Query Parameters
<NR3> in units of Meters/Hertz^2
Range
MPND with units of Meters/Hertz^2
Default Value
NA
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:L2 10.3E-10
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:L2?
10.3E-10
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:L3 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:L3?
Description
This command requires a 4-Port VNA instrument. Sets the L3 coefficient of the match device on the given port of the LRL CALB calibration on the given channel. Outputs the L3 coefficient of the match device on the given port of the LRL CALB calibration on the given channel
Cmd Parameters
<NRf> in Henrys/Hertz^3 (Meters per Hertz cubed)
Query Parameters
<NR3> in Henrys/Hertz^3
Range
MPND
Default Value
0
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:L3 1.32E-25
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:L3?
1.32E-25
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:OFF1 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:OFF1?
Description
This command requires a 4-Port VNA instrument. Sets the offset length coefficient1 of the match device on the given port of the LRL CALB calibration on the given channel. Outputs the offset length coefficient1 of the match device on the given port of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:OFF1 1.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:OFF1?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:OFF2 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:OFF2?
Description
This command requires a 4-Port VNA instrument. Sets the offset length coefficient2 of the match device on the given port of the LRL CALB calibration on the given channel.Outputs the offset length coefficient2 of the match device on the given port of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:OFF2 1.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:OFF2?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:OFF3 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:OFF3?
Description
This command requires a 4-Port VNA instrument. Sets the offset length coefficient3 of the match device on the given port of the LRL CALB calibration on the given channel. Outputs the offset length coefficient3 of the match device on the given port of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:OFF3 1.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:OFF3?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:OFFS?
Description
This command requires a 4-Port VNA instrument. The command sets the offset length of the match device on the given port of the LRL CALB calibration on the given channel. The query outputs the offset length of the match device on the given port of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:OFFS 1.0E0
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:OFFS?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:R <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:R?
Description
This command requires a 4-Port VNA instrument. The command sets the resistance of the match device on the given port of the LRL CALB calibration on the given channel. The query outputs resistance of the match device on the given port of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
MPND
Default
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:R 7.5E1
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:R?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:Z0 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:PORT{1-4}:MATCH:Z0?
Description
This command requires a 4-Port VNA instrument. The command sets the impedance of the match device on the given port of the LRL CALB calibration on the given channel. The query outputs the impedance of the match device on the given port of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
MPND
Default
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:Z0 7.5E1
:SENS1:CORR:COLL:LRL:CALB:DEV1:PORT1:MATCH:Z0?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:TYPe <char1>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:DEVice{1-4}:TYPe?
Description
This command requires a 4-Port VNA instrument. The command sets the type of the given device in the LRL CALB calibration on the given channel. The query outputs the type of the given device in the LRL CALB calibration on the given channel.
Cmd Parameters
<char1> LINE | MATCH | DEVICE1 | DEVICE2
Query Parameters
<char1> LINE | MATCH | DEVICE1 | DEVICE2
Range
NA
Default
LINE
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:DEV1:TYP LINE
:SENS1:CORR:COLL:LRL:CALB:DEV1:TYP?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:FREQuency:BREakpoint <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:FREQuency:BREakpoint?
Description
This command requires a 4-Port VNA instrument. The command sets the breakpoint frequency of the LRL CALB calibration on the given channel. The query outputs the breakpoint frequency of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default
3.000000000E+009
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:FREQ:BRE 1.0E7
:SENS1:CORR:COLL:LRL:CALB:FREQ:BRE?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:OPEN:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:OPEN:OFFS?
Description
This command requires a 4-Port VNA instrument. The command sets the offset length of the open like reflection of the LRL CALB calibration on the given channel. The query outputs the offset length of the open like reflection of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:OPEN:OFFS 1.0E0
:SENS1:CORR:COLL:LRL:CALB:OPEN:OFFS?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:REFPlane <char1>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:REFPlane?
Description
This command requires a 4-Port VNA instrument. The command sets the location of the reference plane in the LRL CALB calibration on the given channel to either the middle or the end of the calibration line. The query outputs the location of the reference plane in the LRL CALB calibration on the given channel.
Cmd Parameters
<char1> MIDdle | END
Query Parameters
<char1> MID | END
Range
NA
Default
END
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:REFP MID
:SENS1:CORR:COLL:LRL:CALB:REFP?
 
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:SHORT:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL:CALB:SHORT:OFFS?
Description
This command requires a 4-Port VNA instrument. The command sets the offset length of the short like reflection of the LRL CALB calibration on the given channel. The query outputs the offset length of the short like reflection of the LRL CALB calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:CALB:SHORT:OFFS 1.0E0
:SENS1:CORR:COLL:LRL:CALB:SHORT:OFFS?