VectorStar™ MS4640A SeriesMicrowave Vector Network Analyzer : SCPI Commands : :SENSe{1-16}:CORRection:COLLect:LRL[:CALa] Subsystem
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa] Subsystem
The :SENSe{1-16}:CORRection:COLLect:LRL[:CALa] subsystem commands provide control of Line-Reflect-Line first calibration (or CALa) configuration parameters, execution, and output reporting. The :CALa keyword is optional for 2-Port VNA instruments and it required for 4-Port VNA instruments.
LRL Calibration Subsystems
The LRL-related calibration commands are organized into five (5) subsystems in the following sections:
:SENSe{1-16}:CORRection:COLLect:LRL:CALB Subsystem
:SENSe{1-16}:CORRection:COLLect:LRL:DEVice{1-4} Subsystem
:SENSe{1-16}:CORRection:COLLect:LRL:PORT Subsystem - 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:LRL:SINGleton Subsystem - 4-Port VNAs
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa] Subsystem (this subsystem)
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND:COUNt <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND:COUNt?
Description
The use of the optional :CALa parameter requires a requires a 4-Port VNA instrument. Sets the number of bands to use in the LRL CALA calibration on the given channel. Outputs the number of bands to use in the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is an integer.
Query Parameters
<NR1> The output parameter is an integer.
Range
1 or 2
Default
1
Syntax Example
:SENS1:CORR:COLL:LRL:BAND:COUN 2
:SENS1:CORR:COLL:LRL:BAND:COUN?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND1:REFLection:TYPe <char1>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND1:REFLection:TYPe?
Description
The use of the optional :CALa parameter requires a requires a 4-Port VNA instrument. Sets the band1 reflection type of the LRL CALa calibration on the given channel. where:
OPENlike = The Band 1 reflection is more like an open.
SHORTlike = The Band 1 reflection is more like a short.
BOTH = The Band 1 reflection is like both an open and a short.
Outputs the band1 reflection type of the LRL CALa calibration on the given channel.
Cmd Parameters
<char1> OPENlike | SHORTlike | BOTH
Query Parameters
<char1> OPEN | SHORT | BOTH
Range
NA
Default
OPEN
Syntax Example
:SENS1:CORR:COLL:LRL:BAND1:REFL:TYP OPEN
:SENS1:CORR:COLL:LRL:BAND1:REFL:TYP?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND2:REFLection:TYPe <char1>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:BAND2:REFLection:TYPe?
Description
The use of the optional :CALa parameter requires a requires a 4-Port VNA instrument. Sets the band2 reflection type of the LRL CALa calibration on the given channel, where:
OPENlike = The Band 2 reflection is more like an open.
SHORTlike = The Band 2 reflection is more like a short.
BOTH = The Band 3 reflection is like both an open and a short.
Outputs the band2 reflection type of the LRL CALA calibration on the given channel.
Cmd Parameters
<char1> OPENlike | SHORTlike | BOTH
Query Parameters
<char1> OPEN | SHORT | BOTH
Range
NA
Default
OPEN
Syntax Example
:SENS1:CORR:COLL:LRL:BAND2:REFL:TYP BOTH
:SENS1:CORR:COLL:LRL:BAND2:REFL:TYP?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:LINE
Description
The use of the optional :CALa parameter requires a 4-Port VNA instrument. Initiates collection of the indicated Device Line Standard data for the CALa calibration on the indicated channel. No query.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:LINE
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:LINE:FREQuency <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:LINE:FREQuency?
Description
The use of the optional :CALa parameter requires a 4-Port VNA instrument. Sets the reference frequency for loss of the given device of the LRL CALa calibration on the given channel. Outputs the reference frequency for loss of the given device of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:LINE:FREQ 1.0E7
:SENS1:CORR:COLL:LRL:DEV1:LINE:FREQ?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:LINE:LENGth <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:LINE:LENGth?
Description
The use of the optional :CALa parameter requires a 4-Port VNA instrument. Sets the line length of the given device of the LRL CALA calibration on the given channel. Outputs the line length of the given device of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:LINE:LENG 1.0E0
:SENS1:CORR:COLL:LRL:DEV1:LINE:LENG?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:LINE:LOSS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:LINE:LOSS?
Description
The use of the optional :CALa parameter requires a 4-Port VNA instrument. Sets the line loss of the given device of the LRL CALa calibration on the given channel. Outputs the line loss of the given device of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in dB/mm.
Query Parameters
<NR3> The output parameter is in dB/mm.
Range
MPND/1000 (MPND divided by 1000)
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:LINE:LOSS 3.0E0
:SENS1:CORR:COLL:LRL:DEV1:LINE:LOSS?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:C0 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:C0?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets capacitance of the match device on the given port of the LRL CALa calibration on the given channel. The C0 (C zero) coefficient is in Farads. Outputs capacitance of the match device on the given port of the LRL CALa calibration on the given channel.
The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument.
Cmd Parameters
<NRf> The input parameter is in Farads.
Query Parameters
<NR3> The output parameter is in Farads.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:C0 3.01E-12
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:C0?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:C1 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:C1?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets the C1 (C one) coefficient of the match device on the given port of the LRL CALA calibration on the given channel. The C1 coefficient is in Farads/Hertz.
Outputs the C1 coefficient of the match device on the given port of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz.
Query Parameters
<NR3> The output parameter is in Farads/Hertz.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:C1 2.0E0
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:C1?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:C2 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:C2?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets the C2 coefficient of the match device on the given port of the LRL CALA calibration on the given channel. The C2 coefficient is in Farads/Hertz^2. Outputs the C2 coefficient of the match device on the given port of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^2.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^2.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:C2 2.0E0
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:C2?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:C3 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:C3?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets the C3 coefficient of the match device on the given port of the LRL CALa calibration on the given channel. The C3 coefficient is in Farads/Hertz^3. Outputs the C3 coefficient of the match device on the given port of the LRL CALA calibration on the given channel
Cmd Parameters
<NRf> The input parameter is in Farads/Hertz^3.
Query Parameters
<NR3> The output parameter is in Farads/Hertz^3.
Range
MPND
Default Value
See Calibration Component Parameters.
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:C3 2.0E0
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:C3?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:L0 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:L0?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets inductance of the match device on the given port of the LRL CALa calibration on the given channel. The L0 (zero) coefficient is in Henrys. Outputs inductance of the match device on the given port of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Henrys.
Query Parameters
<NR3> The output parameter is in Henrys.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:L0 2.0E-6
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:L0?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:L1 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:L1?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets the L1 (L one) coefficient of the match device on the given port of the LRL CALa calibration on the given channel. The L1 coefficient is in Henrys/Hertz. The use of Port 3 and/or Port 4 requires a 4-Port VNA instrument. Outputs the L1 coefficient of the match device on the given port of the LRL CALa calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:L1
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:L1?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:L2 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:L2?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets the L2 coefficient of the match device on the given port of the LRL CALa calibration on the given channel. The L2 coefficient is in Henrys/Hertz^2.
Outputs the L2 coefficient of the match device on the given port of the LRL CALa calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^2.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^2.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:L2 2.0E0
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:L2?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:L3 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:L3?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets the L3 coefficient of the match device on the given port of the LRL CALa calibration on the given channel. The L3 coefficient is in Henrys/Hertz^3. Outputs the L3 coefficient of the match device on the given port of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Henrys/Hertz^3.
Query Parameters
<NR3> The output parameter is in Henrys/Hertz^3.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:L3 2.0E0
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:L3?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:OFF1 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:OFF1?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets the offset Length1 coefficient of the match device on the given port of the LRL CALa calibration on the given channel. The OFF1 coefficient is measured in Meters/Hertz. Outputs the offset Length1 coefficient of the match device on the given port of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz.
Query Parameters
<NR3> The output parameter is in Meters/Hertz.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:OFF1 2.0E0
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:OFF1?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:OFF2 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:OFF2?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets the offset Length2 coefficient of the match device on the given port of the LRL CALa calibration on the given channel. The OFF2 coefficient is measured in Meters/Hertz^2. Outputs the offset Length2 coefficient of the match device on the given port of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^2.
Query Parameters
<NR3> The output parameter is in Meters/Hertz^2.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:OFF2 2.0E0
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:OFF2?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:OFF3 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:OFF3?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets the offset Length3 coefficient of the match device on the given port of the LRL CALA calibration on the given channel. The OFF3 coefficient is measured in Meters/Hertz^3. Outputs the offset Length3 coefficient of the match device on the given port of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters/Hertz^3.
Query Parameters
<NR3> The output parameter is in Meters/Hertz^3.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:OFF3 2.0E0
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:OFF3?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:OFFS?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets offset length of the match device on the given port of the LRL CALA calibration on the given channel. Outputs offset length of the match device on the given port of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:OFFS 1.0E0
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:OFFS?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:R <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:R?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets resistance of the match device on the given port of the LRL CALa calibration on the given channel. Outputs resistance of the match device on the given port of the LRL CALa calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
MPND
Default
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:R 7.5E1
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:R?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:Z0 <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:PORT{1-4}:MATCH:Z0?
Description
The use of the optional :CALa parameter, Port 3, and/or Port 4 require a 4-Port VNA instrument. Sets impedance of the match device on the given port of the LRL CALA calibration on the given channel. Outputs impedance of the match device on the given port of the LRL CALa calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Ohms.
Query Parameters
<NR3> The output parameter is in Ohms.
Range
MPND
Default
5.00000000000E+001
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:Z0 7.5E1
:SENS1:CORR:COLL:LRL:DEV1:PORT1:MATCH:Z0?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:TYPe <char1>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:DEVice{1-4}:TYPe?
Description
The use of the optional :CALa parameter requires a 4-Port VNA instrument. Sets the type of the given device in the LRL CALA calibration on the given channel. Outputs the type of the given device in the LRL CALa calibration on the given channel.
Cmd Parameters
<char1> LINE | MATCH | DEVICE1 | DEVICE2
Query Parameters
<char1> LINE | MATCH | DEVICE1 | DEVICE2
Range
NA
Default
LINE
Syntax Example
:SENS1:CORR:COLL:LRL:DEV1:TYP LINE
:SENS1:CORR:COLL:LRL:DEV1:TYP?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:FREQuency:BREakpoint <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:FREQuency:BREakpoint?
Description
The use of the optional :CALa parameter requires a 4-Port VNA instrument. Sets the breakpoint frequency of the LRL CALA calibration on the given channel. Outputs the breakpoint frequency of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Hertz.
Query Parameters
<NR3> The output parameter is in Hertz.
Range
MPND
Default
3.00000000000E+009
Syntax Example
:SENS1:CORR:COLL:LRL:FREQ:BRE 1.0E7
:SENS1:CORR:COLL:LRL:FREQ:BRE?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:OPEN:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:OPEN:OFFS?
Description
The use of the optional :CALa parameter requires a 4-Port VNA instrument. Sets the offset length of the open like reflection of the LRL CALA calibration on the given channel. Outputs the offset length of the open like reflection of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:OPEN:OFFS 1.0E0
:SENS1:CORR:COLL:LRL:OPEN:OFFS?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:REFPlane <char1>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:REFPlane?
Description
The use of the optional :CALa parameter requires a 4-Port VNA instrument. Sets the location of the reference plane in the LRL CALA calibration on the given channel where:
MIDdle = the middle of the transmission line
END = the end of the transmission line
Outputs the location of the reference plane in the LRL CALA calibration on the given channel.
Cmd Parameters
<char1> MIDdle | END
Query Parameters
<char1> MID | END
Range
NA
Default
END
Syntax Example
:SENS1:CORR:COLL:LRL:REFP MID
:SENS1:CORR:COLL:LRL:REFP?
 
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:SHORT:OFFS <NRf>
:SENSe{1-16}:CORRection:COLLect:LRL[:CALa]:SHORT:OFFS?
Description
The use of the optional :CALa parameter requires a 4-Port VNA instrument. Sets the offset length of the short like reflection of the LRL CALA calibration on the given channel. Outputs the offset length of the short like reflection of the LRL CALA calibration on the given channel.
Cmd Parameters
<NRf> The input parameter is in Meters.
Query Parameters
<NR3> The output parameter is in Meters.
Range
MPND
Default
0.00000000000E+000
Syntax Example
:SENS1:CORR:COLL:LRL:SHORT:OFFS 1.0E0
:SENS1:CORR:COLL:LRL:SHORT:OFFS?