VectorStar™ MS4640A Series Microwave Vector Network Analyzers : Power Menus - 4-Port VNAs : Maximum and Minimum Power Settings
 
Maximum and Minimum Power Settings
For all configurations, the maximum power setting is +15 dBm. For all configurations below 54 GHz, the minimum power setting is –30 dBm and for above 54 GHz, the minimum power setting is –60 dBm. These minimum power settings can be further lowered by application of internal attenuators, linear power calibration effects, and/or the connected presence of OEM millimeter-wave modules.
The power settings are set by the appropriate Start, Stop, and/or Single Power buttons and their related field toolbars.
Below 54 GHz, the user defined start, stop, and/or single power settings are then modified by the effects of any applied attenuators (if equipped), and the relevant Linear Power Calibration and displayed as the appropriate read-only Effective Power button.
Above 54 GHz, the user defined start, stop, and/or single power settings are then modified by the effects of the relevant Linear Power Calibration (noting that internal attenuator effects are not applied above 54 GHz) and displayed as the appropriate read-only Effective Power button.
Depending the VNA model and the equipped test set, where the frequency range as less than 54 GHz, greater than 54 GHz, or across 54 GHz, some of the following read-only Effective Power buttons appear:
Effective Start Power
Effective Start Power < 54 GHz – Button is only available when 3743A or MA25300A Modules are in use. If not present, only the Effective Start Power button present.
Effective Start Power > 54 GHz – Button is only available when 3743A or MA25300A Modules are in use. If not present, only the Effective Start Power button present.
Effective End Power
Effective Stop Power < 54 GHz – Button is only available when 3743A or MA25300A Modules are in use. If not present, only the Effective End Power button present.
Effective Stop Power > 54 GHz – Button is only available when 3743A or MA25300A Modules are in use. If not present, only the Effective End Power button present.
Effective Single Power
Effective Single Power < 54 GHz – Button is only available when 3743A or MA25300A Modules are in use. If not present, only the Effective Single Power button present.
Effective Single Power > 54 GHz – Button is only available when 3743A or MA25300A Modules are in use. If not present, only the Effective Single Power button present.
For VNAs where frequencies either cross 54 GHz (MS4647A VNAs) or use frequencies above 54 GHz (MS4647A VNAs with Option 080/081 or MS4640A VNAs with Option 082/082 and E-Band or W-Band modules), additional buttons of Effective Start Power > 54 GHz, Effective Stop Power > 54 GHz, and Effective Single Power > 54 GHz are present.
Where compatible with the various options above, the 4-Port VNA power maximum and power minimum are the same as for the 2-Port VNA less the loss of the Test Set and cables.
A summary of the power settings is in Table: Summary of Maximum and Minimum Power Levels.
 
Summary of Maximum and Minimum Power Levels 
VNA, Power Level, and Frequency Range
Power Sweep Type
Power
Setting
Max/Min1
Factors Affecting Effective Power
Resultant Effective Power
Any MS4640A VNA
Maximum Power Above, Below, or Across 54 GHz
Any
+15 dBm maximum
Option 061/062 Attenuators2
Linear Power Calibration
Effects at left can lower effective maximum power.
Any MS4640A VNA
Minimum Power for
Less Than 54 GHz
Power Sweep CW
Single Power Mode
Power Sweep
Swept Frequency
Power Sweep
Swept Frequency Single Power
–30 dBm minimum
Option 061/062 Attenuators
Linear Power Calibration
Effects at left can lower effective minimum power to less than -30 dBm.
MS4647A VNAs with BB/mm-Wave Option 080/0813
Minimum Power for Less Than
54 GHz
to
Greater Than 54 GHz
Power Sweep
Swept Frequency
Power Sweep
Swept Frequency
Single Power
–60 dBm minimum
Below 54 GHz:
Option 061/062 Attenuators
Linear Power Calibration
The default minimum power for frequencies less than 54 GHz is
-30 dBm and for frequencies greater than 54 GHz is -60 dBm.
The effects at left are applied to these default values can lower the effective minimum power below the default value in each frequency range.
Above 54 GHz:
Attenuators have no effect above 54 GHz.
Linear Power Calibration
MS4640A VNAs with mm-Wave Option 082/083 and OML/VDI mm-Wave Modules4
Any
–100 dBm minimum
VNA Attenuators have no effect on OML/VDI modules at any frequency.
Linear Power Calibration
The default minimum power for any frequency is always -100 dBm.
The effects at left are applied to these default values can lower the effective minimum power below the default value in each frequency range.

1 The default power setting applies to the user configurable Start, Stop, and Single Power buttons and their variants.

2 The Active Measurement Suite - Option 061 provides two attenuators; one if the Source 1 Path, and one in the Receive 2 Path. Option 062 provides four attenuators; One in each Source Path and one in each Receive Path.

3 Broadband/Millimeter-Wave System – ME7838A or ME7838D system with MS4647A VNA, Option 080/081, 3739x Test Set, and 3743A or MA25300A Modules to 145 GHz.

4 Millimeter-Wave System – Any MS4640A VNA, with Option 082/083, 3739x Test Set, and OML or VDI Millimeter-Wave Modules.