The CAL OPTIONS menu and related sub-menus provide access to advanced calibration options either before or after a calibration procedure has been completed.
• MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS
CAL OPTIONS (CALIBRATION OPTIONS) Menu
Flexible Cal Status (Off/On)
Read only display button. Flexible calibration is the ability to select a sub-set of a currently active calibration status. For example, if a Full 2 Port calibration has been completed, flexible calibration allows the calibration to be reduced to a Transmission Frequency Response calibration. Select toggles flexible calibration off and on.
Flexible Cal Setup
Use Flexible Cal Setup to change the parameters above. Select displays the FLEXIBLE CAL SETUP dialog box.
Interpolation allows additional interpolated measurement points between calibrated measurement points. This is useful if the user wants to zoom into a specific area without having to recalibrate the instrument. The interpolated points must lie within the calibration frequency points. Select toggles interpolation OFF and ON with a default state of OFF.
Apply Isolation (Off/On)
If this button is unavailable, isolation calibration was not performed. The isolation calibration procedure is started by a button on the calibration type menu.
If this button is available, isolation calibration was performed, and select toggles isolation calibration OFF and ON.
Perform Manual Adapter Removal
Adapter removal permits accurate measurement of non-insertable devices using an adapter of known electrical length and two full 12-term calibrations. Manual adapter removal extracts the behavior of the adapter from the setup after a successful calibration. Select displays the MANUAL ADAPTER REMOVAL dialog box.
Cal Merge allows merging two calibration files into a single file where the calibrations can use different methods at different frequencies. The calibrations must be of the same calibration type. Select displays the CAL MERGE dialog box.
Hybrid calibrations take two 1-port calibrations and hybridizes them into a 2-port calibration which allows a mixed-media calibration for a device that may have coaxial and waveguide connections. Select displays the HYBRID CAL menu where the hybrid calibration is configured.
Use network extraction to generate an S-Parameter (.s2p) file for a set of networks. The file can ben be embedded or de-embedded as required. Select displays the NETWORK EXTRACTION dialog box.
The flexible calibration setup feature makes it possible to select a subset of the S-Parameters to be corrected from the available full-term calibration.
Procedure
1. Perform the necessary full-term calibration.
2. On the CAL OPTIONS menu, select Flexible Cal Setup.
• MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Flexible Cal Setup
• The FLEXIBLE CAL SETUP dialog box appears.
3. Select the necessary S-Parameters in any combination as long as one S-Parameter is selected:
• S11
• S12
• S21
• S22
4. Select the Apply Selection button to apply the calibration.
5. Select Close to close the dialog box and return to the CAL OPTIONS menu.
6. If Full Term Cal (By Port) is selected above, the area label is Port Selections (Full Term) with available options of:
• Port 1 and/or
• Port 2
7. If Reflection Only (By Port) is selected above, the area label is Port Selections (Reflection Only) with the available options of:
• Port 1 and/or
• Port 2
8. If Customize Cal (By S-Param) is selected, the S-Param Selections area appears with S-Parameter selection check boxes. Select any combination of:
• S11
• S21
• S12
• S22
9. Select Apply Selection to apply the selections.
10. Select Close to close the dialog without saving the selections.
MANUAL ADAPTER REMOVAL Dialog Box - 2-Port VNAs
Use manual adapter removal to extract the electrical behavior of an adapter after a successful calibration procedure. This is especially useful when the DUT configuration is not entirely compatible with common calibration procedures such as having different connectors at each end.
• MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Perform Manual Adapter Removal | MANUAL ADAPTER REMOVAL Dialog Box
MANUAL ADAPTER REMOVAL Dialog Box
Instructions
Adapter removal permits accurate measurement of non-insertable devices. The process involves using an adapter of known electrical length and performing two full 12-term calibrations. In the procedure below, the Y file is the file with the calibration when the adapter connected to port 1 and the X file is the file with the calibration when the adapter connected to port 2.
Procedure
1. Connect the adapter to port X, where X signifies any port. Perform a full 12-term (Full 2 Port) calibration using Y' and Y as the test ports and store the calibration to disk.
2. Connect the adapter to port Y, where Y signifies any port that is not X. Perform a full 12-term calibration using X and X' as the test ports and store calibration to disk.
3. Call up the X and Y files.
4. Input the estimated adapter electrical length.
5. Select Perform Adapter Removal to remove adapter.
CAL MERGE Dialog Box - 2-Port VNAs
Calibration merge allows merging two calibration files into a single file where the prior calibrations can use different methods (such as SOLT and SSLT) at different frequencies. The calibrations must be the same calibration type (such as Full 1 Port) and be available on the instrument hard drive. Calibration merge has other requirements (described below) as to the number of points, start and stop frequencies, and front panel settings.
• MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Cal Merge | CAL MERGE Dialog Box
CAL MERGE Dialog Box
Instructions
Calibration merge (cal merge) allows the merging of two RF calibration files into a single file. The RF calibrations can use different methods and different frequencies.
Requirements and Notes
• The two RF calibrations must be of the same cal type (such as Full 2 Port) and must be stored on the VectorStar VNA hard drive.
• The combined frequency list for the two source cal files cannot exceed the maximum allowable instrument points, which is either 25,000 or 100,000 points depending on MAX POINTS menu selection.
• The frequency lists and matching correction terms are combined. For the frequencies which coincide, the terms from the first cal file are used.
• The start and stop frequencies are adjusted to include the entire frequency range provided by the two cal files.
• The first cal file provides all other front panel setup configuration information.
• Since Cal Merge can result in a non-discrete step size, the cal merge file sweep type is set to a frequency-based segmented sweep.
Procedure
1. In the dialog box, select Browse to select the appropriate cal files (e.g. calfile1.chx and calfile2.chx).
2. Click the First File or Second File Browse button to display an OPEN (Setup.cal File) dialog box.
3. Select Perform Cal Merge to perform the merge of the frequency lists and matching correction terms.
4. If required, enter the Cutoff Frequency in kHz.
5. Click Perform Cal Merge to merge the calibration files.
6. If the selected files are incorrect, a warning message is displayed.
7. Best practices recommend saving the resultant Cal Merge file, but a save is not required.
• MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Hybrid Cal | HYBRID CAL
CAL OPTIONS (CALIBRATION OPTIONS) Menu
Hybrid Cal Setup
Hybrid calibrations take two 1-port calibrations and hybridizes them into a 2-port calibration which allows a mixed-media calibration for a device that may have coaxial and waveguide connections. It also allows 2-port calibrations where the two 1-port calibrations are not restricted to be the same type. Select displays the HYBRID CAL SETUP dialog box where the hybrid calibration is configured.
After the hybridized calibration has been configured in the HYBRID CAL SETUP dialog box, make sure all through connections are complete. When ready to proceed, select the Thru button to start the hybridization. The display dims for a few seconds, and then reappears with a completion check box.
Done
If the Done button is unavailable, the hybrid calibration has not been successfully completed. After a successful completion, the Done button is available. Select returns to the CALIBRATION menu where the Cal Status button is set to ON.
• MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Hybrid Cal | HYBRID CAL | Hybrid Cal Setup | HYBRID CAL SETUP Dialog Box
HYBRID CAL SETUP Dialog Box - 2-Port VNAs
Instructions
A hybrid calibration takes two 1-port calibrations and hybridizes them into a 2-port calibration. This allows a mixed-media calibration for a device that may have both coaxial and waveguide connections. It also allows 2-port calibrations where the two 1-port calibrations are not restricted to the same type such as.
Hybridizing Two 1-Port Cals into a 2-Port Cal
In the File 1 field, select the Browse button to navigate to the appropriate Channel Setup and Calibration (CHX) file for the first calibration.
Repeat for the File 2 field.
Thru/Recip Selection
The figure shows the relationship between the two ports and the Thru 1-2 Info button is automatically selected.
OK
When both files have been identified, select OK to return to the HYBRID CAL menu.
NETWORK EXTRACTION Dialog Box - 2-Port VNAs
The network extraction features provides a method of generating an S-Parameter (.s2p) file for a set of networks. The.s2p file can then be embedded or de-embedded into the error coefficient of the VNA as required. Four extractable network configurations are provided:
• Type A – Adapter Extraction – Extract one 2-port network
• Type B – Two Tier Calibration – Extract one 2-port network
• Type C – Inner and Outer Calibrations Available – Extract two 2-port networks
• Type D – Outer Cal Only – Extract two 2-port networks.
• MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Network Extraction | NETWORK EXTRACTION Dialog Box
NETWORK EXTRACTION Dialog Box
NETWORK EXTRACTION Dialog Box – Choose the type of desired extraction from the four (4) buttons below:
1. Type A - Extract one (1) 2-Port Network - Adapter Extraction
2. Type B - Extract one (1) 2-Port Network - Two Tier Calibration
3. Type C - Extract two (2) 2-Port Networks - Inner and Outer Cals Available
4. Type D - Extract two (2) 2-Port Networks - Outer Cal Only using divided-by-2 method
5. Legend for Dialog Box Figures:
• a – Location of one or more Reference Planes for Cal A
• b – Location of one or more Reference Planes for Cal B
• # – Locations “n” indicated Cal Type for n-Port
Instructions
The network extraction feature provides the method for generating the S-Parameter for a given set of networks. This generated.s2p file can then be embedded or de-embedded into the error coefficient for the VNA at a later stage. The dialog box provides four extraction choices accessible via large buttons. Each choice provides a separate configuration dialog box as described in the sections below.
• Extract One 2-Port Network - Type A - Adapter Extraction
• MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Network Extraction | NETWORK EXTRACTION Dialog Box | Extract Two 2-Port Networks - Type A - Adapter Extraction
NETWORK EXTRACTION Dialog Box - Type A - Adapter Extraction
Use Type A Network Extraction to extract one (1) 2-Port network using adapter extraction.
Description
Network Extraction provides the means of generating SnP files of networks. The generated files can then be embedded or de-embedded. Based on the type of extraction chosen, multiple SnP files may be generated, as shown in the graphics for each extraction type. Port Swapping can be performed in the Embedding/De-embedding menus.
Requirements
All calibration files must be:
• Full 12 term calibrations
• Of the same type (e.g. all SOLT/SOLR)
• Over the exact same frequency points
Instructions
1. Use the Browse button to select the Cal A file. Repeat to select the Cal B file. If necessary, estimate the Port 1 Length in ps.
2. Select Perform Network Extraction to perform the extraction.
3. If the extraction is successful, follow the prompt to save the generated SnP files/s.
4. After the extraction, the focus auto-returns to the CAL OPTIONS menu.
• MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Network Extraction | NETWORK EXTRACTION Dialog Box | Extract Two 2-Port Networks - Type B - Two Tier Calibration
NETWORK EXTRACTION Dialog Box - Type B - Two Tier Calibration
Use Type B Network Extraction to extract one (1) 2-Port network using available two tier calibration.
Description
Network Extraction provides the means of generating SnP files of networks. The generated files can then be embedded or de-embedded. Based on the type of extraction chosen, multiple SnP files may be generated, as shown in the graphics for each extraction type. Port Swapping can be performed in the Embedding/De-embedding menus.
Requirements
• Calibration A must share a common test port.
• Calibration B in this extraction type must only be a full 1-port calibration, which is ideal if a Through is not available.
• Both cals have to be of the same Cal Type and use the exact same frequency points.
Instructions
1. Use the Browse button to select the Cal A file. Repeat to select the Cal B file. If necessary, estimate the Port 1 Length in ps.
2. Select Perform Network Extraction to perform the extraction.
3. If the extraction is successful, follow the prompt to save the generated SnP files/s.
4. After the extraction, the focus auto-returns to the CAL OPTIONS menu.
• MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Network Extraction | NETWORK EXTRACTION Dialog Box | Extract Two 2-Port Networks - Type C - Outer Cal Only
NETWORK EXTRACTION Dialog Box - Type C - Inner and Outer Cals
Use Type C Network Extraction to extract two (2) 2-Port networks using available Inner and Outer calibrations.
Description
Network Extraction provides the means of generating SnP files of networks. The generated files can then be embedded or de-embedded. Based on the type of extraction chosen, multiple SnP files may be generated, as shown in the graphics for each extraction type. Port Swapping can be performed in the Embedding/De-embedding menus. All calibration files must be:
• Full 12-term calibrations
• The same calibration type
• Over the exact same frequency points
Instructions:
1. Use the Browse button to select the Cal A file. Repeat to select the Cal B file.
2. Select Perform Network Extraction to perform the extraction.
3. If the extraction is successful, follow the prompt to save the generated SnP files/s.
4. After the extraction, the focus auto-returns to the CAL OPTIONS menu.
• MAIN | Calibration | CALIBRATION | Cal Options | CAL OPTIONS | Network Extraction | NETWORK EXTRACTION Dialog Box | Extract Two 2-Port Networks - Type D - Outer Cal Only
NETWORK EXTRACTION Dialog Box - Type D - Outer Cal Only Divide-By-2 Method
Use Network Extraction to extract two (2) 2-Port networks using Outer calibrations and the divide-by-2 method.
Description
Network Extraction provides the means of generating SnP files of networks. The generated files can than be embedded or de-embedded. Based on the type of extraction chosen, multiple SnP files may be generated, as shown in the graphics for each extraction type. Port Swapping can be performed in the Embedding/De-embedding menus.
Application
These extraction types are for cases where an inner-cal is not possible. The network measured is assumed symmetrical, and SnP files generated using Divide by 2 schemes. An option is given to zero-out the match terms instead of fully allocating them to the outer-ports.
Instructions
1. Make sure the appropriate calibration is active.
2. Zero-out the match terms if needed.
3. Connect the network and select Perform Network Extraction to perform the extraction.
4. If the extraction is successful, follow the prompt to save the generated SnP files/s.
5. After the extraction, the focus auto-returns to the CAL OPTIONS menu.
The USER DEFINED MATCH DEVICES dialog box only appears if Device 2 or Device 4 is set to Match and the user clicks the Match Info button.
USER DEFINED MATCH DEVICES Dialog Box
Microstrip Info Button
Select displays the MICROSTRIP INFO Dialog Box shown below. In this example, the 25 Mil Kit was selected.
MICROSTRIP INFO Dialog Box for SSLT and 25 Mil Kit
Test Port 2 Connector Area - Standard Info Button
Select displays the Standard Info (SOLT/R) K-Conn (M) Dialog Box shown below. In the example below, K-Conn (M) was selected for a Test Port 1 DUT Connector.
Test Port 2 Connector Standard Info Dialog Box for SOLT/R and K-Conn (M)
The Standard Info dialog shows the applicable calibration parameters for Test Port 2.
Test Port 1 Connector Area - Standard Info Button
Select displays the STANDARD INFO (SOLT/R) STANDARD LABEL (V-Conn M) Dialog Box shown below. Note that the name of this dialog changes depending on the selected Cal Method and DUT Connector. In the example below, V-Conn (M) was selected for a Test Port 1 DUT Connector.
Note
The name of this dialog and the information presented is dependent on the user selections for Calibration Type, Calibration Method, and Test Port Connectors.
Test Port 1 Connector Standard Info Dialog Box for SOLT/R V-Conn (M)
The Standard Info dialog shows the applicable calibration parameters for Test Port 1.
CAL KIT/AUTOCAL Menu - 2-Port VNAs
Use the CAL KIT/AUTOCAL menu to install, save, and restore calibration kit characterization files between an external memory device, the instrument firmware, and a hard drive on the instrument or on a network.
• MAIN | Calibration | CALIBRATION | Cal Kit/AutoCal Characterization | CAL KIT/AUTOCAL
CAL KIT/AUTOCAL Menu - 2-Port VNAs
Install Kit/Charac.
Select loads the Calibration Kit file or AutoCal Characterization file from the hard drive or external memory device into the VNA firmware through the INSTALL (AutoCal Characterization/Cal Kit File) dialog box.
Select saves the Cal Kit or AutoCal Characterization file from the firmware to the location of choice (typically the instrument hard drive) for later use through the SAVE (AutoCal Characterization/Cal Kit) File dialog box.
Use the INSTALL (AUTOCAL CHARACTERIZATION/CAL KIT) dialog box to install the calibration kit coefficients file or AutoCal Characterization files in the instrument firmware for subsequent use.A recommended best practice is to keep the cal kit serial number as part of the file name.
1. In the Select File Type area, select either the AutoCal Characterization or the Cal Kit radio button.
2. In the Open field, either enter the file name or click Browse to navigate manually.
3. If AutoCal Characterization was selected, the file type will be an AutoCal Characterization ACD file.
4. If Cal Kit was selected, the file type will be a Cal Kit Coefficient.ccf file.
5. Click Open to load the file or Cancel to return to the menu.
SAVE (AUTOCAL CHARAC./CAL KIT) Dialog Box - 2-Port VNAs
Use the SAVE (AUTOCAL CHARACTERIZATION/CAL KIT) dialog to save a Calibration Kit Coefficients file or an AutoCal Characterization file from the VNA firmware to external location such as the instrument hard drive, a network drive, or an external memory device. This is useful for storing in the instrument hard drive multiple files from available cal kits or AutoCal modules.
An alternate method is to use a Windows program such as File Manager to copy the file from the supplied USB memory device onto the hard drive. In that case, we recommend using the default locations mentioned later in this paragraph.
• MAIN | Calibration | CALIBRATION | Cal Kit/AutoCal Characterization | CAL KIT/AUTOCAL | Save Kit/Charac | SAVE (AUTOCAL CHARACTERIZATION/CAL KIT) Dialog Box
SAVE (AUTOCAL CHARACTERIZATION/CAL KIT FILE Dialog Box
1. AutoCal Characterization radio button selected at left. 2. Cal Kit radio buttons selected at right.
Save Dialog Button and Field Availability
The available buttons and fields on the dialog depend on the radio button selection in the Select File Type area.
• If AutoCal Characterization is selected, no other buttons/fields are available and instructions are immediately below.
• If Cal Kit is selected, the For Cal Kit Only area becomes available with fields for Line Type, Cal Method, and Cal Kit Name. Instructions are in the next sub-section following.
Instructions for AutoCal Characterization File Type
1. Click OK to save the AutoCal Characterization file.
• Click Cancel to stop the save.
2. The SAVE AS (AUTOCAL CHARACTERIZATION ACD FILE) dialog box appears.
3. Navigate to the required storage location:
• C:\AnritsuVNA\AutoCal is recommended.
4. Click Save. The system auto-returns to the Cal Kit/AutoCal menu.
• MAIN | Calibration | CALIBRATION | Cal Kit/AutoCal Characterization | CAL KIT/AUTOCAL | Display/Edit Info | CAL KIT INFO Dialog Box
CAL KIT INFO Dialog Box
1. Controls for AutoCal Characterization files at left. 2. Controls for Cal Kit files at right.
Button and Field Availability
The available buttons and fields on the dialog depend on the radio button selection in the Select File Type area.
• If AutoCal Characterization is selected, no other buttons/fields are available.
• If Cal Kit is selected, the For Cal Kit Only area becomes available with fields for Line Type, Cal Method, and Select Cal Standard.
Instructions
1. Select AutoCal Characterization or Cal Kit as required.
2. If Cal Kit was selected above, in the For Cal Kit Only area, select a Line Type value:
• Coaxial
• Non-Dispersive
• Waveguide
3. Select a Cal Method value:
• SOLT/SOLR
• SSLT
• SSST
4. Select a Cal Standard value:
• 0.8 mm (M)
• 0.8 mm (F)
• W1-Conn (M)
• W1-Conn (F)
• V-Conn (M)
• V-Conn (F)
• K-Conn (M)
• K-Conn (F)
• 2.4 mm (M)
• 2.4 mm (F)
• GPC-3.5 (M)
• GPC-3.5 (F)
• SMA (M)
• SMA (F)
• N-Conn (M)
• N-Conn (F)
• N-Conn (75) (M)
• N-Conn (75) (F)
• GPC-7
• 7/16 (M)
• 7/16 (F)
• TNC (M)
• TNC (F)
• User-Defined1 (M) through User-Defined32 (M)
• User-Defined1 (F) through User-Defined32 (F)
5. Click Display/Edit.
• The STANDARD INFO read-only dialog box appears.
• Note that the dialog box title and content fields are dependent on the selections made in the steps above for Cal Method, Line Type, and Connector Type.
• For example, if:
• Calibration type is set to Full 2 Port (12 Term Cal)
• Calibration method is set to SOLT/SOLR
• Line type is set to Coaxial
• Connector type is set to V-Conn (M).
• The read-only dialog box is titled:
• STANDARD INFO (SOLT/R) STANDARD LABEL (V-Conn (M)) (shown below).
Standard Info (SOLT/R) Standard Label (V-Conn (M)) Dialog Box
6. Click OK to close the read-only dialog box.
RESTORE DEFAULT COEF. Dialog Box - 2-Port VNAs
Use the RESTORE DEFAULT COEF. dialog box to restore firmware-stored Cal Kit Coefficients fields back to their default coefficients. For best performance, either install the cal kit coefficients file supplied with your cal kit, or enter your user-defined coefficients before starting this procedure. The restore function is not available to AutoCal kits as they do not have restorable characterization data.
Use this dialog to restore factory coefficients to available calibration kits.
1. Select the Line Type as required.
2. Select the Calibration Kits as required to be restored.
3. Click OK.
Available Selections
The table below shows the available calibration kits in the Select Cal Kit field of the RESTORE DEFAULT COEFFICIENTS dialog box. The available kits depend on the input combination selected for Line Type Media and Cal Method
Calibration Kit Availability in the RESTORE DEFAULT COEF. Dialog Box
LINE TYPE Media Setting
CAL METHOD Setting
Available Calibration Kits
Coaxial
SOLT/SOLR
0.8 mm Conn, W1-Conn, V-Conn, K-Conn, 2.4 mm, 2.4 mm V, GPC-3.5, SMA, N-Conn, N-Conn (75), GPC-7, 7/16, TNC
SSLT
0.8 mm Conn, W1-Conn
SSST
0.8 mm Conn, W1-Conn
Non-Dispersive
SOLT/SOLR
0.8 mm Conn, W1-Conn, V-Conn, K-Conn, 2.4 mm, GPC-3.5, SMA, N-Conn, N-Conn (75), GPC-7, 7/16, TNC