Power Meter : Power Meter : Frequency Menu with Offset Function
  
Frequency Menu with Offset Function
Key Sequence: Frequency
A user defined frequency offset can be entered to adjust the frequency displayed on the instrument from the actual swept frequency. When enabled Offset will be displayed at the bottom of the screen (Figure: 200 MHz Frequency Offset Example) and the Center Freq, Start Freq, and Stop Freq keys will indicate that a frequency offset has been turned on.
Set the Freq Offset to 0 Hz to remove the frequency offset.
Note 
The Freq Offset will affect the displayed values of Frequencies and Limits. The currently frequency offset value is displayed in the Freq 2/2 Menu.
Power Meter Freq 1/2 with Offset Function Menu
(Offset) Center Freq
Sets the frequency at the center of the measurement. Frequencies can be entered in units of GHz, MHz, kHz, or Hz. Enter the frequency by using the keypad, the rotary knob, or the arrow keys. When the center frequency is entered, the labeling below the analog display shows the center frequency and the span in the most appropriate units based upon the value.
(Offset) Start Freq
Enter a start frequency by using the keypad, the rotary knob, or the arrow keys. If the entered start frequency is greater than the current stop frequency, then the stop frequency is automatically adjusted to be 1 kHz greater than the start frequency (Min Span).
(Offset) Stop Freq
Enter a stop frequency by using the keypad, the rotary knob, or the arrow keys. If the entered stop frequency is lower than the current start frequency, then the start frequency is automatically adjusted to be 1 kHz less than the stop frequency (Min Span).
Span
Opens the Span Menu.
Signal Standard
Opens the signal standard menu, showing the currently selected signal standard and a complete list of signal standards to choose from.
Channel #
Sets the channel information for the selected standard. If the particular standard has not been used before, then the channel number defaults to the lowest legal channel number for that standard. If that standard has been used before, then the last used channel will be the default.
Full Band
Sets the frequency of the unit to a 100 MHz span on the current center frequency.
Step Size & Offset
Opens the Freq 2/2 Menu.
Freq 2/2 Menu
Key Sequence: Freq > Step Size & Offset
Power Meter Freq 2/2 with Offset Function Menu
Freq Step
Sets the interval that is used by the arrow keys. Enter the step amount by using the keypad, the rotary knob, or the arrow keys.
Freq Offset
Enter the desired offset (positive or negative) using the keypad, the arrow keys, or the rotary knob. If entering a frequency using the keypad, the submenu key labels change to GHz, MHz, kHz, and Hz. Press the appropriate units key. Pressing the Enter key has the same affect as the MHz submenu key.
Offset Step Size
Enter the desired frequency offset step size. The offset frequency step specifies the amount by which the offset frequency will change when the Up/Down arrow keys are pressed.
Use the keypad or the rotary knob to change the Offset Step Size.
Back
 