Vector Network Analyzer Measurement Guide : VNA View Menus : Calibration Menus : Cal Kit Definition Menus for Coax
 
Cal Kit Definition Menus for Coax
Cal Kit Definition Menus for Coax
 
DUT Connector Selector List Box for Coax
DUT Connector Selector List Box for Cable
This list box (Figure: DUT Connector Selector List Box for Cable) is opened by the Coax DUT Port # soft key. In this example, the Cal Type is Full 2 Port, and the Cal Method is SOLT. Note that the Cal Kit data includes capacitance coefficients for the Open and inductance coefficients for the Short. Inductance values are provided in all instruments with VNA mode.
This feature is used by all Anritsu VNA Masters and by the S412E LMR Master.
 
 
SOLT – Coax Menu
Key Sequence: Shift, Calibrate (2) > DUT Port Setup > Setup User‑Defined > Enter
SOLT – Coax Menu
 
 
Edit Name
Press this soft key to use the Text Entry feature (refer to the Text Entry section in the user guide for your instrument). Enter a name for your setup.
Open
Press this soft key to display the Open menu. Enter any offset length and capacitance values. See Figure: Cal Kit Definition Menus for Coax.
Short
Press this soft key to display the Short menu. Enter any offset length and inductance values. See Figure: Cal Kit Definition Menus for Coax.



Back
Press this soft key to return to the DUT Port Setup Menu (Coax).
 
 
DUT Port Setup Menu (Waveguide)
Key Sequence: Shift, Calibrate (2) > DUT Port Setup
DUT Port Setup Menu
WG DUT Port 1
Press this soft key to open the “DUT Connector Selector” List Box for Port 1. Choose the connector type that matches the DUT connection on Port 1. This indicates that you are using the matching calibration kit, and the VNA Master displays the corresponding calibration coefficients that will be used during the calibration procedure.
Examples:
WG11A
WG12
WG20
SSLT(1) – USER 1
SSLT(4) – USER 4
WG DUT Port 2
Press this soft key to open the “DUT Connector Selector” List Box for Port 2. Choose the connector type that matches the DUT connection on Port 2. This indicates that you are using the matching calibration kit, and the VNA Master displays the corresponding calibration coefficients that will be used during the calibration procedure.
Setup User-Defined
Press this soft key to open the “User‑Defined Cal Kit Selector” List Box (Figure: DUT Connector Selector List Box for Waveguide). Format: SSST(#) – USER #. Selecting a user‑defined cal kit displays the cal kit definition menu (Figure: Cal Kit Definition Menus for Waveguide). In this menu, you may enter the calibration coefficients for their specific calibration kit.
Back
Press this soft key to return to the Calibration Menu.