VNA Master™Programming Manual : VNA Commands : [:SENSe]:CORRection:CKIT Subsystem
 
[:SENSe]:CORRection:CKIT Subsystem
This subsystem provides commands that modify and configure the device under test (DUT).
[:SENSe]:CORRection:CKIT Subsystem 
Keyword
Parameter Data or Units
[:SENSe]
 
:CORRection
 
:CKIT
 
:USER{1-4}
 
Calibration Connector Information
[:SENSe]:CORRection:CKIT:INFormation?
Description
Returns a string of information of the given calibration connector. <connector> defines the connector family and can be given in either long or short form. [connector-name] is a string that defines the name that is associated with the given <connector> and is optional. [connector-name] must be enclosed by parentheses. Note that the connector must be valid for the current calibration line type.
Note that user 1, user 2, user 3, or user 4 is based on the current calibration method.
The query response begins with an ASCII header. The header specifies the number of following bytes. It appears in the format #AX, where A is the number of digits in X, and X is the number of bytes that follow the header. Parameters are returned in comma-delimited ASCII format. Each parameter is returned as “NAME=VALUE[UNITS]”.
The tables below list the available connectors and connector names that are associated with the calibration line types. Note that Coax User cal kits do not have a calibration name associated with them.
COAX
<connector>
[connector-name]
Description
NMALe
OSLN50
N-Conn(M)
Cal Kit: OSLN50.
If no connector-name is given, information will default to this.
Query Response: <block> (returns comma-delimted ASCII format)
NMALe
SLN50A or OSLN50A-8 or OSLN50A-18
N-Conn(M)
Cal Kit: OSLN50A-8 or OSLN50A-18
Query Response: <block> (returns comma-delimted ASCII format)
NMALe
TOSLN50A or TOSLN50A-8 or TOSLN50A-18
N-Conn(M)
Cal Kit: TOSLN50A-8 or TOSLN50A-18
Query Response: <block> (returns comma-delimted ASCII format)
NFEMale
OSLNF50
N-Conn(F)
Cal Kit: OSLNF50.
If no connector name is given, then information will default to this.
Query Response: <block> (returns comma-delimted ASCII format)
NFEMale
OSLNF50A or OSLNF50A-8 or OSLNF50A-18
N-Conn(F)
Cal Kit: OSLNF50A-8 or OSLNF50A-18
Query Response: <block> (returns comma-delimted ASCII format)
NFEMale
TOSLNF50A or TOSLNF50A-8 or TOSLNF50A-18
N-Conn(F)
Cal Kit: TOSLNF50A-8 or TOSLNF50A-18
Query Response: <block> (returns comma-delimted ASCII format)
KMALe
OSLK50
K-Conn(M)
Cal Kit: OSLK50.
If no connector name is given, then information will default to this.
Query Response: <block> (returns comma-delimted ASCII format)
KMALe
TOSLK50A or TOSLK50A-20
K-Conn(M)
Cal Kit: TOSLK50A-20
Query Response: <block> (returns comma-delimted ASCII format)
KFEMale
OSLKF50
K-Conn(F)
Cal Kit: OSLKF50
If no connector-name is given, then information will default to this.
Query Response: <block> (returns comma-delimted ASCII format)
KFEMale
TOSLKF50A or TOSLKF50A-20
K-Conn(F)
Cal Kit: TOSLKF50A-20
Query Response: <block> (returns comma-delimted ASCII format)
716Male
2000-767
7/16(M)
Cal Kit: 2000-767.
If no connector-name is given, then information will default to this.
Query Response: <block> (returns comma-delimted ASCII format)
716Male
2000-1618 or 2000-1618-R
7/16(M)
Cal Kit: 2000-1618-R
Query Response: <block> (returns comma-delimted ASCII format)
716Female
2000-768
7/16(F)
Cal Kit: 2000-768
If no connector name is given, then information will default to this.
Query Response: <block> (returns comma-delimted ASCII format)
716Female
2000-1619 or 2000-1619-R
7/16(F)
Cal Kit: 2000-1619-R
Query Response: <block> (returns comma-delimted ASCII format)
TNCMale
 
TNC(M)
Cal Kit: 1091-5x & 1015-55.
If no connector-name is given, then information will default to this.
Query Response: <block> (returns comma-delimted ASCII format)
TNCFemale
 
TNC(F)
Cal Kit: 1091-5x & 1015-54
If no connector-name is given, then information will default to this.
Query Response: <block> (returns comma-delimted ASCII format)
SMAMale
3650
SMA(M)
Cal Kit: 3650
SMAFemale
3650
SMA(F)
Cal Kit: 3650
USR1
 
User 1 cal Kit information for the current calibration method.
Query Response: <block> (returns comma-delimted ASCII format)
USR2
 
User 1 cal Kit information for the current calibration method.
Query Response: <block> (returns comma-delimted ASCII format)
USR3
 
User 1 cal Kit information for the current calibration method.
Query Response: <block> (returns comma-delimted ASCII format)
USR4
 
User 1 cal Kit information for the current calibration method.
Query Response: <block> (returns comma-delimted ASCII format)
Syntax
[:SENSe]:CORRection:CKIT:INFormation? <connector>, [connector-name string]
Cmd Parameter
NA (query only)
Query Parameter
<char> <connector>, [connector-name string]
Query Response
<block> (returns comma‑delimted ASCII format)
Example
To get information for K‑Conn(M) with cal Kit TOSLK50A‑20:
:SENS:CORR:CKIT:INF? KMAL, "TOSLK50A”
or
:SENS:CORR:CKIT:INF? KMAL, "TOSLK50A-20"
The response for this query when using SOLT COAX:
#6000164OPEN=5.010 mm,SHORT=5.010 mm,
THRU=16.070 mm,C0=4.500 e-15,C1=395.000 e-27,
C2=-20.000 e-36,C3=0.400 e-45,L0=4.000 e-12,
L1=-650.000 e-24,L3=39.000 e-33,L3=-0.640 e-42
Front Panel Access
Shift-2 (Calibrate), DUT Port Setup, DUT Port 1